EMC evaluation in radiation-emitting medical devices: regulatory limits and an integrated safety approach.
Federica Censi, Simona D'Agostino, Francesca Apolinio, Edoardo Piero Masini, Giancarlo Burriesci
Department of Cardiovascular, Endocrine-Metabolic Diseases and Aging, Italian National Institute of Health, Rome, Italy. · Department of Occupational and Environmental Medicine, Epidemiology and Hygiene, Italian National Institute for Insurance Against Accidents at Work, Rome, Italy.
Key points
From the paper's conclusion- By combining EMC testing according to IEC 60,601-1-2 with exposure assessment based on IEC 62,311 and comparison with ICNIRP limits, manufacturers can demonstrate full regulatory compliance in terms of EMC.
Abstract
INTRODUCTION: Electromagnetic compatibility (EMC) is a fundamental safety requirement for medical devices, ensuring both the limitation of electromagnetic emissions and adequate immunity to external electromagnetic disturbances. AREAS COVERED: Within the European regulatory framework, EMC is addressed in the Regulation 2017/745 (MDR), which requires manufacturers to minimize radiation emissions, prevent interference with other devices, and provide appropriate information in the instructions for use. The main technical reference standard for EMC is IEC 60,601-1-2, which defines test methods and limits for electromagnetic emissions and immunity. However, this standard evaluates radiated emissions only above 30 MHz, creating a regulatory gap for devices operating at lower frequencies. Moreover, it does not address the assessment of unintended or spurious emissions below this threshold. EXPERT OPINION: To address these limitations, additional standards and guidelines must be considered. IEC 62,311 provides methodologies for assessing human exposure to electromagnetic fields across the frequency range 0 Hz-300 GHz, while exposure limits can be derived from the ICNIRP guidelines and the European Directive 2013/35/EU. By combining EMC testing according to IEC 60,601-1-2 with exposure assessment based on IEC 62,311 and comparison with ICNIRP limits, manufacturers can demonstrate full regulatory compliance in terms of EMC.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Medical Device Electromagnetic Immunity Testing for Exposure to Consumer Wireless Power Transfer Systems - Part II: Test Method.IEEE electromagnetic compatibility magazine · 2026
- Medical Device Electromagnetic Immunity Testing for Exposure to Consumer Wireless Power Transfer Systems - Part I: Emission Characteristics.IEEE electromagnetic compatibility magazine · 2026
- Electromagnetic Compatibility in Electric Vehicles: A Review and Near-Field Case Study of an Electric BusVehicles · 2026
- Occupational radiation exposure and radiation protection practices in interventional radiology and cardiology: A multicenter study in FranceDiagnostic and Interventional Imaging · 2026
- AI medical device post-market surveillance regulations: consensus recommendations by the European Society of RadiologyInsights into Imaging · 2025
- RF-EMF response modeling of implantable medical electrodes under 1.5T and 3T MRI exposure using a lumped element model approach.Physics in medicine and biology · 2026