Safety and tolerability study of subcutaneous injections of larger volumes in abdominal and lower extremity sites suitable for autoinjectors.
Marc C Torjman, Sarah Spiegel, Channy Loeum, Jennifer Lessin, Jeffrey I Joseph
Department of Anesthesiology and Perioperative Medicine, Sidney Kimmel Medical College, Thomas Jefferson University, Philadelphia, PA, USA. · Department of Anesthesiology and Perioperative Medicine, Thomas Jefferson University Hospital, Philadelphia, PA, USA.
Key points
From the paper's conclusion- These results provide evidence that subcutaneous administration of >3 mL volumes into the subcutaneous tissue of the abdomen or thigh, was safe and well tolerated.
Abstract
OBJECTIVES: This study investigated pain and safety of subcutaneous injections, using larger volumes (>3 mL) and injection rates. METHODS: Thirty‑two subjects were recruited and had 4 injections: 2.5, 5.0, 10.0, and 15 mL administered in the abdomen and thigh using an infusion pump. Injection force was measured and pain assessments were obtained using VAS (0-100 mm) and verbal pain ratings. RESULTS: All 125 infusions appropriately delivered the complete volume. VAS pain increased after the start of injections: median VAS: 10.00 mm (IQR: 1.50-27.00) and peaked to 16 mm (IQR: 8.50-43.50) at 4 min for the 15 mL infusions. Median VAS after termination of infusions was 6.5 mm (IQR: 0-18). Verbal pain at start of injection was none to mild. VAS pain ratings were higher (p < 0.001) for 0.500 mL/sec and 0.333 mL/sec infusion rates compared to 0.023 mL/sec. Peak injection forces were significantly (p < 0.001) different from one another for the three infusion rates with a moderate correlation (R = 0.55, p < 0.001) between injection force and pain. Minimal saline leaked onto the skin surface for all volumes and delivery rates. CONCLUSION: These results provide evidence that subcutaneous administration of >3 mL volumes into the subcutaneous tissue of the abdomen or thigh was safe and well tolerated. CLINICAL TRIAL REGISTRATION: NCT06194578.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Adapting Drug Delivery for High-Volume Home Administration: Poster Presented at PDA Week 2026.PDA journal of pharmaceutical science and technology · 2026
- Usability of a High-Volume Autoinjector (HVAI): Addressing the User Need for Rapid High-Volume Subcutaneous Drug Delivery: Poster Presented at PDA Week 2026.PDA journal of pharmaceutical science and technology · 2026
- Volume-dependent detection of free intraperitoneal fluid by FAST: a prospective in-vivo study with sequential fluid instillation.European journal of trauma and emergency surgery : official publication of the European Trauma Society · 2026
- Advancing Subcutaneous Delivery: Evaluation of Near-Body Auto-Injector Systems for High-Volume Biologic Administration with Human Hyaluronidase: Poster Presented at PDA Week 2026.PDA journal of pharmaceutical science and technology · 2026
- Effectiveness of High-Volume Corticosteroid Injection Diluted with Saline in Subacromial Impingement Syndrome: A Randomized Trial.Indian journal of orthopaedics · 2026
- Patient-Centric Development of a Low-Volume, Low-Viscosity, Once Every 4 Weeks (Q4W) Autoinjector Combination Product for PovetaciceptbioRxiv · 2026