Design and biomechanical evaluation of a patient-specific 3D-printed wrist splint for carpal tunnel syndrome: a finite element analysis study.
Desi Wang
Department of Medical Engineering, The 954th Army Hospital of the Chinese People's Liberation Army, Xizang, China.
Key points
From the paper's conclusion- This single-volunteer proof-of-concept study supports the feasibility of a lightweight, anatomy-matched splint with mechanically validated global displacement under static loading. Effects on ventilation, skin comfort, and CTS symptoms require physiological and clinical evaluation.
Abstract
INTRODUCTION: Conventional wrist splints for carpal tunnel syndrome (CTS) may have limited anatomical fit and ventilation. This study developed and mechanically evaluated an anatomy-matched, 3D-printed wrist splint using stress-guided Voronoi lattice optimisation. MATERIALS AND METHODS: Wrist and forearm geometry from one healthy adult volunteer was acquired by structured-light scanning (0.2 mm point distance). A 3.0 mm shell was optimised by preserving longitudinal and transverse load-bearing features and removing material from low-strain regions. A coupled splint-soft-tissue finite element analysis (FEA) model with frictional contact (μ = 0.4) simulated a static 50 N wrist-flexion load. Global displacement predictions were assessed by mechanical bench testing of the printed polylactic acid (PLA) splint. A non-porous 3.0 mm solid shell derived from the same anatomy served as the design comparator. RESULTS: Under the simulated load, maximum von Mises stress was 18.4 MPa against a PLA yield strength of 60 MPa, with a minimum safety factor of 3.26. Predicted distal displacement was 2.15 mm, compared with 2.32 mm in bench testing (7.33% relative error). Compared with the solid-shell control, the optimised design reduced mass by 45.93% and achieved a 45.2% open surface area while retaining adequate static load-bearing capacity. CONCLUSIONS: This single-volunteer proof-of-concept study supports the feasibility of a lightweight, anatomy-matched splint with mechanically validated global displacement under static loading. Effects on ventilation, skin comfort, and CTS symptoms require physiological and clinical evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Design and Finite Element Evaluation of a Customized 3D-Printed Wrist-to-Finger Orthosis Targeting Carpal Tunnel Pressure Reduction.Polymers · 2026
- Surrogate model-assisted numerical analysis of biomechanical properties for carpal tunnel syndrome.Clinical biomechanics (Bristol, Avon) · 2026
- Modular Cage-Type Prosthetic Construction for Lumbar Intervertebral Reconstruction: A Computer-Aided Design-Based Geometric Design and Finite Element Analysis Study.Cureus · 2026
- Stiffness and Strength Enhancement of Variable-Thickness Triply Periodic Minimal Surface Shell Lattices via Stress-Feedback-Driven Design.Materials (Basel, Switzerland) · 2026
- Computational analysis of Lattice-Integrated transfemoral prosthetic socketsMaterials & Design · 2026
- Feasibility of Generatively Designed Patient-Specific Implants for Orthognathic Surgery: An In Silico Comparative Study.International dental journal · 2026