Low-cost TiO2 passive sensor for detection of glyphosate in airborne particles: laboratory and field evaluation.
Alejandra R Vences, Alejandra Castro-Carranza, Iván Y Hernández-Paniagua, Orlando Hernández-Cristóbal, Jocelyn Olivares-Salazar, Gema L Andraca-Ayala, Aron Jazcilevich-Diamant, Jürgen Gutowski
International Laboratory of Environmental Electron Devices (LAIDEA UNAM), ENES Campus Morelia, National Autonomous University of Mexico UNAM, Morelia, Mexico. · Institute of Atmospheric Sciences and Climate Change (ICAyCC), University City, National Autonomous University of Mexico UNAM, Mexico City, Mexico.
Key points
From the paper's conclusion- These results support TiO2 passive surfaces to be suitable as low-cost, field-deployable option for monitoring airborne pesticides, and show their potential to be integrated in sensing electronic devices.Highlights• Detection of airborne glyphosate through low-cost TiO2 passive sensors.• High surface affinity of TiO2 enabling near-instant adsorption of glyphosate.• HPLC-FLD results in accordance w
Abstract
This study evaluates low-cost sol-gel titanium dioxide (TiO2) films as passive sensing surfaces for detecting airborne glyphosate (Gly), leveraging its adsorption capabilities. TiO2 film samples were exposed in the laboratory to analytical grade and commercial Gly (1 and 10 µg/ml) by spin and spray coating, and were deployed for 15 days in an avocado-growing area of Michoacán, Mexico, where Gly is routinely applied. Samples were analysed by scanning electron microscopy (SEM) with energy-dispersive X-ray spectroscopy (EDS), and UV-vis spectrophotometry, and the results were validated by high-performance liquid chromatography coupled with fluorescence detection (HPLC-FLD). SEM-EDS revealed Gly aggregates on the TiO2 surface, with C, N and P localized over the oxide. Gly adsorption attenuated the TiO₂ absorbance peak at about 300 nm, increasingly with rising concentration of pure Gly. HPLC-FLD confirmed Gly to be present in all samples. The field-exposed sensor's peak area was 1.9 times that of the field blank and 73% of that of a 360 µg/ml laboratory reference. These results support TiO2 passive surfaces to be suitable as low-cost, field-deployable option for monitoring airborne pesticides, and show their potential to be integrated in sensing electronic devices.Highlights• Detection of airborne glyphosate through low-cost TiO2 passive sensors.• High surface affinity of TiO2 enabling near-instant adsorption of glyphosate.• HPLC-FLD results in accordance with UV-vis spectroscopy and SEM findings.• Scalable methodology for glyphosate monitoring in agricultural regions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Label-free glyphosate biosensor based on tilted fiber Bragg grating with gold nanostars.Optics express · 2026
- Development of an o-COSAN Ion-Pair Complex-Modified PVC Membrane for Microconductometric Glyphosate Detection.Micromachines · 2026
- Sampling and SERS detection of pesticides in air and water using a bifunctional PTFE membrane functionalized with high-index Au trisoctahedrons and ZIF-67.Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy · 2026
- Zinc Oxide/Ag Water-Gated Field-Effect Transistor Coupled with Surface-Enhanced Raman Scattering for Indigo Dye Detection.ACS omega · 2026
- Plasmonic Nanoink-Coated Paper SERS Substrate for Pesticide DetectionACS Applied Nano Materials · 2025
- Smartphone-based visual quantitation of glyphosate in liquid media using effervescent colorimetric tablets encompassing Ru-doped MOF with augmented enzyme-like activity.Mikrochimica acta · 2026