Scandinavian journal of clinical and laboratory investigation

Long-term storage stability of uracil in plasma.

Malene H Kirkeby, Elke Hoffmann-Lücke, Eva Greibe

Department of Clinical Biochemistry, Aarhus University Hospital, Aarhus N, Denmark.

Key points

From the paper's conclusion
  • In conclusion, no systematic change in P-uracil was observed during storage at -80 °C for up to three years; however, variability between repeated measurements was observed at individual level.

Abstract

5-fluorouracil (5-FU) is widely used in cancer treatment. Both 5-FU and uracil are metabolized by dihydropyrimidine dehydrogenase (DPD); therefore, the European Medicines Agency (EMA) recommends measuring plasma uracil (P-uracil) and/or DPYD genotyping before treatment initiation. EMA classifies partial DPD deficiency as P-uracil ≥ 16  µg/L and <150 µg/L and complete deficiency as P-uracil ≥ 150  µg/L, why accurate measurement is crucial. This study investigated the three-year stability of P-uracil at -80 °C storage in seventy-eight patient samples collected during routine work-up before 5-FU initiation in 2022. Following initial analysis, samples were stored at -80 °C and reanalyzed in December 2025. P-uracil was measured using an in-house LC-MS/MS method. Seventy-six patients were included in the statistical analysis. Median P-uracil was 15.7 µg/L µg/L (range 6.2-69.0 µg/L) in 2022 and 15.0 µg/L (7.0-98.7 µg/L) in 2025, with no significant systematic difference (p = 0.241). The median percentage difference (PD%) was -2.2% (range -55.1% to +80.1%). A PD% exceeding ±15% was observed in 34 samples across all P-uracil levels: 53.3%, 50.0%, and 36.8% of samples with P-uracil <10, 10 to <16, and ≥16 µg/L, respectively. Samples with P-uracil ≥16 µg/L showed the widest PD% range (-55.1% to 80.1%). In nine patients, variability led to crossing the clinical threshold of 16 µg/L (absolute P-uracil change of 3.43-11.45  µg/L). In conclusion, no systematic change in P-uracil was observed during storage at -80 °C for up to three years; however, variability between repeated measurements was observed at individual level.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Long-term storage stability of uracil in plasma — Science Skim