In-line HCl/NaClO decontamination for carryover control in piezoelectric droplet-printing digital PCR

Zhaonan Liu, Hongwei Quan, Mengqi Kong, Zonghao Li, Yuntong Gao, Deding Tang, Yan Gong, Jie Hu

Key points

From the paper's conclusion
  • Here, we evaluated four nozzle-cleaning strategies under fixed printing conditions: a de-ionized-water rinse, two commercial nucleic acid decontamination reagents (NH9901 and VNR11-01), and an automated in-line HCl/NaClO tip-cleaning procedure (G4).

Abstract

Piezoelectric droplet-on-demand printing enables controlled partition generation for digital polymerase chain reaction (dPCR) and spatially addressable droplet arrays. However, a shared dispensing nozzle introduces a risk of sample-to-sample carryover. Here, we evaluated four nozzle-cleaning strategies under fixed printing conditions: a de-ionized-water rinse, two commercial nucleic acid decontamination reagents (NH9901 and VNR11-01), and an automated in-line HCl/NaClO tip-cleaning procedure (G4). Cleaning efficacy was assessed by qPCR for carryover and inhibition and independently cross-validated by direct dPCR partition counting using two sample types. Under single-cycle conditions, G4 showed no detectable carryover: no-template control (NTC) amplification within 45 qPCR cycles and no positive dPCR partitions, without measurable inhibition in the corresponding positive controls. Zero positives among N = 19 469–20 412 analyzed NTC partitions correspond to a one-sided 95% upper bound of ≈3/N (≈1.5 × 10−4 per partition); these are reported as below the detection limit. Droplet-generation characteristics were preserved after a single cleaning cycle, and the digital signal response was proportional (FAM R2 = 0.9998; ROX R2 = 0.9972) with 3.31% within-session repeatability (n = 20). A 20-cycle assessment at a nominal 3.33 × 102 copies/μL kept NTC concentrations below the detection limit at all evaluated checkpoints with positive control (PC) concentrations of 3.62 × 102–3.64 × 102 copies/μL. Exploratory analysis of 493 clinical specimens assessed workflow applicability in complex biological matrices without diagnostic performance evaluation. This in-line decontamination strategy provides a practical approach for contamination-controlled nozzle operation in piezoelectric droplet-printing dPCR under the evaluated single-nozzle and single-platform conditions; broader robustness and long-term material compatibility remain to be established.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

In-line HCl/NaClO decontamination for carryover control in piezoelectric droplet-printing digital PCR — Science Skim