A vacuum ultraviolet spectrometer for impurity, relative electron temperature and Lyman opacity monitoring in MAST-U divertor.
V A Soukhanovskii, A G McLean, T Macwan, M Robson, I Coffey, J Macdonald, E Rose, R Scannell,
Lawrence Livermore National Laboratory, Livermore, California 94550, USA. · United Kingdom Atomic Energy Authority, Culham Centre for Fusion Energy, Abingdon, Oxfordshire, United Kingdom.
Key points
From the paper's conclusion- To monitor plasma impurities, relative electron temperature, and deuterium Lyman series opacity in advanced divertor configurations, a lower divertor-viewing vacuum ultraviolet vacuum ultraviolet (VUV) spectrometer survey, poor resolution, extended domain with a fast camera detector was deployed on the mega-ampere spherical tokamak upgrade tokamak.
Abstract
To monitor plasma impurities, relative electron temperature, and deuterium Lyman series opacity in advanced divertor configurations, a lower divertor-viewing vacuum ultraviolet vacuum ultraviolet (VUV) spectrometer survey, poor resolution, extended domain with a fast camera detector was deployed on the mega-ampere spherical tokamak upgrade tokamak. Operating the diagnostic on the medium-size spherical tokamak with plasma currents up to 1 MA and NBI heating up to 3.6 MW introduced new technical and operational challenges, yet enabled new measurements relevant to future fusion devices and essential to understanding and modeling divertor radiative processes. Spectral data were collected using a commercial McPherson Inc. Model 251 spectrograph, equipped with a micro-channel plate image intensifier and configured for a 250-1650 Å wavelength range. Key challenges included (1) optimizing the viewing geometry to avoid VUV reflections from graphite plasma-facing components; (2) developing a differentially pumped ultra-high vacuum system to sustain up to a 104 deuterium pressure difference. Two fast, up to 0.5 kHz, visible detectors were tested: a CCD camera optically coupled with a 4 m fiberoptic imaging bundle and a one-dimensional NMOS camera coupled via a fiberoptic faceplate. The spectrometer recorded deuterium Lyman series and low-Z impurity spectra in a variety of divertor configurations, including the Super-X, the X-point target, and the double-decker divertors. Initial data suggested that carbon line intensity ratios could monitor relative electron temperatures between and during edge localized modes. This approach relies on Δn = 0, 1, 2 line intensity sensitivities within the 1-30 eV range; however, its effectiveness depends on average divertor temperatures exceeding the current level of a few eV. Relative ratios of Lyβ to Dα intensities were used to monitor deuterium opacity and will serve to benchmark 2D plasma and radiation transport modeling.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Interpretation of the Balmer series-limit recombination spectrum in the MAST-U divertor.The Review of scientific instruments · 2026
- High-resolution visible spectroscopy for ion temperature and flow velocity measurements of the TCV divertor plasmaReview of Scientific Instruments · 2025
- Spectral irradiance and power responsivity calibrations from 200 nm to 400 nm using the new NIST Ultraviolet Spectral Comparator FacilityMetrologia · 2026
- Synchronized operation of new, spectrally filtered X-point fast camera with multiple fast cameras to study filamentary transport in MAST-U.The Review of scientific instruments · 2026
- The OMEGA-extended performance streaked optical pyrometer with spectral resolution.The Review of scientific instruments · 2026
- Development and calibration of a broadband extreme ultraviolet-vacuum ultraviolet spectrometer.The Review of scientific instruments · 2026