Conceptual design of Thomson scattering system for detecting electron temperature anisotropy in magnetically confined plasma.
Kodai Hoshi, Hisamichi Funaba, Kentaro Sakai, Hiyori Uehara, Masato Ota, Haruki Kawaguchi, Hiroyuki Yamaguchi, Ryo Yasuhara
The Graduate University for Advanced Studies, SOKENDAI, Toki, Gifu 509-5292, Japan. · National Institute for Fusion Science, Toki, Gifu 509-5292, Japan.
Key points
From the paper's conclusion- These results indicate that the proposed configurations provide a basis for time-resolved measurements of electron temperature anisotropy in CHD.
Abstract
In magnetically confined plasmas, electron temperature anisotropy can be formed as a difference between the electron temperatures parallel and perpendicular to the magnetic field. In Compact Helical Device (CHD), a high temporal resolution Thomson scattering system is planned to be installed, which may enable detailed measurements of the temporal evolution of anisotropy. In this study, we designed Thomson scattering system configurations for detecting electron temperature anisotropy in CHD and evaluated the feasibility of anisotropy detection. We found that the single-pass configuration using an additional observation system can detect a temperature difference of 1.4 times between the perpendicular and parallel directions in the core region, although its detection performance deteriorates in the peripheral region. In contrast, the double-pass configuration using an additional counter-propagating laser path can detect a difference of 1.3 times over a wide radial range. These results indicate that the proposed configurations provide a basis for time-resolved measurements of electron temperature anisotropy in CHD.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Implementation of the Thomson scattering system for the SMall Aspect Ratio Tokamak (SMART): Status and synthetic modeling.The Review of scientific instruments · 2026
- Temporal weight-position mapping: time-resolved parameter dominance of electron dynamics in nonlinear Thomson scattering.Optics express · 2026
- Thomson Scattering Measurement of 1 keV Electron Temperature on Lawson Machine 26arXiv · 2026
- Thomson scattering in high-background plasmas using polarization-resolved spectroscopy.The Review of scientific instruments · 2026
- Generalized theory of second-harmonic orthogonally polarized dispersion interferometry and implementation for pulsed power electrode plasma measurements.The Review of scientific instruments · 2026
- Numerical evaluation of Paschen-α Zeeman spectroscopy for recombination front position measurement in JT-60SA detached divertor plasmaNuclear Fusion · 2026