The Review of scientific instruments

Assessment of the two-filter soft x-ray method to infer electron temperature in the sub-100 eV regime.

M W Lee, Soobin Lim, Ji Hwan Kim, G W Nam, Jongmin Lee, Chan-Hwi Jeon, Y S Hwang, C Sung

Korea Advanced Institute of Science and Technology, Daejeon 34141, Republic of Korea. · Q-Beam Solution Inc., 160 Daehwa-ro, Daedeok-gu, Daejeon, Republic of Korea.

Key points

From the paper's conclusion
  • Overall, the results indicate that the various random and systematic effects must be properly accounted for to achieve two-filter Te measurements in the sub-100 eV regime.

Abstract

The two-filter soft x-ray (SXR) method has been widely used to infer the electron temperature (Te) in keV-scale fusion plasmas, but its application in the sub-100 eV regime has been relatively less explored. For application in this regime, several additional effects require consideration. This work assesses the two-filter SXR method to infer Te in the sub-100 eV regime using the Be/Al filter configuration of the versatile experiment spherical torus SXR arrays. In this process, the relevant effects are separated into two categories. Random uncertainty components, namely, the filter-thickness fabrication tolerance and detector noise, are propagated through the ratio-temperature response curve to quantify the uncertainty. Systematic effects, namely, native oxide layers on the filters and impurity line emission, modify the response and are incorporated into the response calculation. Overall, the results indicate that the various random and systematic effects must be properly accounted for to achieve two-filter Te measurements in the sub-100 eV regime.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Assessment of the two-filter soft x-ray method to infer electron temperature in the sub-100 eV regime — Science Skim