The Review of scientific instruments

Lock-in infrared thermography: Phase analysis for rapid, wide-range thermal conductivity measurements.

Ethan A Scott, Jeffrey L Braun, Jessica Reyes, Bruce Bolliger, Terrence Soares, John T Gaskins, Marko J Tadjer, Patrick E Hopkins

Department of Mechanical and Aerospace Engineering, University of Virginia, Charlottesville, Virginia 22904, USA. · Laser Thermal Analysis, Inc., 937 2nd St SE, Charlottesville, Virginia 22902, USA.

Review

Key points

From the paper's conclusion
  • We report on a phase-based lock-in thermography approach, combined with a multilayered thermal model (often employed in thermoreflectance analysis), to measure the thermal conductivity of a variety of materials.

Abstract

We report on a phase-based lock-in thermography approach, combined with a multilayered thermal model (often employed in thermoreflectance analysis), to measure the thermal conductivity of a variety of materials. The spatial distribution of the material's thermal phase is monitored with an infrared camera, which is locked into the frequency of a modulated laser used to heat the material. This phase distribution is then fit with a thermal model, in which properties such as thermal conductivity are extracted as fit parameters. This approach enables non-contact, front-side measurements, which are insensitive to surface roughness. The technique does not strictly require the application of a transducer layer, but we highlight the practical benefits of applying a removable adhesive layer to serve as a near-surface absorber. We demonstrate the efficacy of the method by measuring materials with thermal conductivities that span more than three orders of magnitude (∼1 to >2000 W m-1 K-1).

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Lock-in infrared thermography: Phase analysis for rapid, wide-range thermal conductivity measurements — Science Skim