Design and validation of a heated shear cell for in situ x-ray scattering under controlled temperature and deformation.
Alex M Jordan, Shuquan Cui, Frank S Bates, Christopher J Ellison
Department of Engineering & Technology, University of Wisconsin-Stout, Menomonie, Wisconsin 54751, USA. · Department of Chemical Engineering and Materials Science, University of Minnesota, Minneapolis, Minnesota 55455, USA.
Key points
From the paper's conclusion- We report the design and validation of a compact heated shear cell for in situ structural characterization of materials under combined temperature and deformation control.
Abstract
We report the design and validation of a compact heated shear cell for in situ structural characterization of materials under combined temperature and deformation control. The instrument applies prescribed linear reciprocating displacement to a parallel-plate geometry using programmable logic controller-based open-loop control, enabling reproducible shear deformation across experimentally relevant temperatures, amplitudes, and cycle frequencies. Emphasis is placed on quantitative validation of instrument performance, including measurements of temperature accuracy, temporal stability, plate-to-plate uniformity, and displacement repeatability during continuous shear at elevated temperatures. Independent temperature control of the stationary and moving plates ensures uniform thermal conditions at the sample, while the mechanical architecture accommodates large displacement amplitudes without introducing rotational or parasitic motion. An open Kapton window provides unobstructed access to the sample region, allowing simultaneous shear, heating, and real-time x-ray scattering measurements. The validated performance demonstrates reliable and stable operation under conditions relevant to soft matter and polymer processing studies. Representative in situ small-angle x-ray scattering measurements acquired at a synchrotron beamline further confirm stable, low-background data acquisition under coupled thermomechanical conditions, validating the instrument for its intended application. The modular design and straightforward control architecture establish the shear cell as a robust and adaptable platform for coupled mechanical deformation and in situ structural measurements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- SIRIUS-DIMBOTER: A Synchrotron In Situ Research Instrument for Uniaxial Stretching in Diverse Immersion Media across a Broad Temperature Range.Analytical chemistry · 2026
- Integrated laser heating stage with active geometry modulation for simultaneous in situ x-ray transmission and evolved gas analysis of molten liquids.The Review of scientific instruments · 2026
- Measuring structural restraint in an amorphous material using the variance of speckle intensityarXiv · 2026
- A high-pressure sample cell for small-angle x-ray scattering.The Review of scientific instruments · 2026
- On the delay-and-sum processing of shear wave signals from a scanning thermo-elastic laser ultrasound system.The Journal of the Acoustical Society of America · 2026
- Soft material mechanical property degradation at high strain ratesarXiv · 2026