Development of a high-field pulsed magnet and optical fiber coupled cryostat system for magneto-photoluminescence measurements.

Deepesh Kalauni, Kingshuk Mukhuti, Tao Peng, Bhavtosh Bansal

原始摘要(英文原文)· Original abstract
This paper presents the development of a high field pulsed magnet system combined with an optical fiber-coupled low temperature cryostat for magneto-photoluminescence measurements. A novel aspect of our system is the use of electrolytic capacitors in a 75 kJ capacitor bank to drive the pulsed magnet, enabling the generation of high magnetic fields at a relatively low charging voltage (400 V). The zylon-reinforced wire wound magnet coil achieves maximum field strength of 35 T with a magnetic field rise time of 10 ms. The integrated cryostat was developed to meet the specific requirements of magneto-photoluminescence experiments using a 3.8 K helium closed-cycle cryocooler to provide a stable low-temperature environment down to 5 K.