Device-Specific Cut-Offs of Canon Aplio i800 2D Shear Wave Elastography for Advanced Fibrosis in MASLD: A Histologically Validated Study.
Wolfgang Kratzer, Andreas Binzberger, Stefan G Kauschke, Finn Dolze, Markus Werner, Thomas F E Barth, Madeleine Feller, Heike Neubauer,
University Hospital Ulm, Department of Internal Medicine I, Germany, Ulm. · Boehringer Ingelheim Pharma GmbH & Co. KG, Department of Cardiometabolic Research, Germany, Biberach.
Key points
From the paper's conclusion- Canon Aplio i800 2D-SWE provides robust and clinically applicable, device-specific non-invasive detection of advanced fibrosis in MASLD and supports implementation into guideline-based stepwise risk stratification.
Abstract
PURPOSE: To evaluate the diagnostic accuracy of Canon Aplio i800 two-dimensional shear wave elastography (2D-SWE) for detecting advanced fibrosis (≥ F3) in metabolically associated steatotic liver disease (MASLD) using histology as reference, and to compare its performance with the FIB-4 index. MATERIALS AND METHODS: In this prospective single-center study, 49 patients with MASLD and available liver histology were included. Fibrosis staging (F0-F4) was assessed by three blinded pathologists according to Brown and Kleiner. Shear wave velocity (m/s) served as the primary elastography parameter. Diagnostic performance for ≥ F3 was assessed by ROC analysis, and clinically applicable rule-in and rule-out thresholds were determined using predefined sensitivity- and specificity-based criteria. RESULTS: Advanced fibrosis (≥ F3) was present in 15/49 patients (30.6%). 2D-SWE demonstrated an AUROC of 0.88 (95% CI 0.78-0.96) for ≥ F3, compared with 0.84 (95% CI 0.71-0.96) for FIB-4 (p = 0.39). A threshold of 1.45 m/s (6.3 kPa) achieved 100% sensitivity and negative predictive value. A threshold of 1.85 m/s (11.0 kPa) yielded 94% specificity and 82% positive predictive value. CONCLUSION: Canon Aplio i800 2D-SWE provides robust and clinically applicable, device-specific non-invasive detection of advanced fibrosis in MASLD and supports implementation into guideline-based stepwise risk stratification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Endoscopic ultrasound strain elastography in MASLD: Evaluation against transient elastography and non-invasive markers.Hepatology forum · 2026
- Non-Invasive Fibrosis Testing in Adults With Obesity and MASLD: A Systematic Review and Meta-Analysis.Diabetes, obesity & metabolism · 2026
- Diabetes and obesity reduce FIB-4 accuracy in MASLD referral pathwaysJHEP Reports · 2026
- Assessment of Non-Invasive Tests for MASLD, MASH and At-Risk MASH in a Multicentre Prospective Real-World Study.Diabetes, obesity & metabolism · 2026
- Noninvasive Detection of Clinically Significant Fibrosis in Biopsy-Proven MASLD: A Retrospective Cross-Sectional Diagnostic Accuracy Study.Medicina (Kaunas, Lithuania) · 2026
- Meta-Analysis: Enhanced Liver Fibrosis (ELF) Test for Identifying Significant Fibrosis, Advanced Fibrosis, and Cirrhosis in Patients With Metabolic Dysfunction-Associated Steatotic Liver Disease.Alimentary pharmacology & therapeutics · 2026