Hydraulic Resistance of the Internal Shunt of a Double-Balloon Catheter for Percutaneous Hepatic Perfusion: An In Vitro Study

Jan Turra, Oliver Biecker, Osman Oecal, Maik Von der Forst, Max Dietrich, Philippe Grieshaber, Folker Wenzel, Jessica Hassel,

Preclinical

Key points

From the paper's conclusion
  • The results are limited to one catheter in a rigid-tube model.

Abstract

Background: Percutaneous hepatic perfusion (PHP) isolates the liver circulation with a double-balloon catheter whose internal shunt maintains venous return from the lower body during balloon occlusion of the inferior vena cava (IVC). Clinically, PHP is associated with hypotension and high catecholamine requirements. This in vitro study quantified the hydraulic resistance of the shunt. Methods: One 16 F double-balloon catheter was tested in a bench model (12.7 mm PVC tubing, open reservoir) using a blood analog (43% glycerol by mass) at 21.4–22.4°C. A centrifugal pump generated flows at 500–4000 RPM with uninflated and inflated balloons (4 mL NaCl); flow and pressure were recorded (n = 5 per step). Results: With uninflated balloons, flow reached 7.55 ± 0.05 L/min at a pressure gradient of 31.6 ± 0.5 mmHg (4000 RPM). After inflation, flow was restricted to the shunt lumen and reduced by 96.9% (500 RPM) to 93.3% (4000 RPM). The maximum shunt flow was 0.51 ± 0.01 L/min at 387 mmHg; at 10 mmHg it was approximately 0.02 L/min. The resistance (ΔP/flow) rose from 493 to 762 mmHg·min/L (p < 0.001). Flow was laminar (Reynolds number < 1800). Conclusion: In this bench model, the internal shunt was a severe hydraulic bottleneck that is unlikely to carry physiological venous return at low venous driving pressures. Occlusion time should be kept short with close hemodynamic monitoring, and shunt design optimization should be investigated. The results are limited to one catheter in a rigid-tube model.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Hydraulic Resistance of the Internal Shunt of a Double-Balloon Catheter for Percutaneous Hepatic Perfusion: An In Vitro Study — Science Skim