RSC advances

Forensic electrochemistry on a drop: bromocresol green-modified screen-printed platforms for dual colorimetric-electrochemical detection of scopolamine.

Larissa M A Melo, Robert D Crapnell, Elena Bernalte, Daniel S Paula, Rodrigo A A Munoz, Wendell K T Coltro, Wallans T P Dos Santos, Craig E Banks

Faculty of Science and Engineering, Manchester Metropolitan University Dalton Building, Chester Street M1 5GD UK c.banks@mmu.ac.uk. · Institute of Chemistry, Federal University of Goiás Campus Samambaia 74690-900 Goiânia Goiás Brazil.

Key points

From the paper's conclusion
  • Differential pulse voltammetry provided a wide linear range from 5 to 80 µg mL-1, with limits of detection (3σ) and quantification (10σ) of 0.02 and 0.07 µg mL-1, respectively, alongside excellent repeatability (RSD < 7.0% for peak current and <2.0% for peak potential).

Abstract

A novel bulk-modified screen-printed electrode (SPE) was developed by incorporating bromocresol green (BG) as a pH-responsive colorimetric reagent, directly into the graphite ink formulation to create an integrated dual-response sensing platform. This bulk modification strategy enables complementary electrochemical and colorimetric responses on a single disposable device (BG-SPE), eliminating the need for additional reagents or multi-step procedures. The platform was applied to the detection of scopolamine, a clinically relevant tropane alkaloid associated with both therapeutic use and drug-facilitated criminal activity. The BG-SPE sensor enabled rapid, portable preliminary visual screening followed by electrochemical confirmation and quantification of scopolamine in beverages and biological matrices. Differential pulse voltammetry provided a wide linear range from 5 to 80 µg mL-1, with limits of detection (3σ) and quantification (10σ) of 0.02 and 0.07 µg mL-1, respectively, alongside excellent repeatability (RSD < 7.0% for peak current and <2.0% for peak potential). Method validation in real beverage samples (whiskey) and synthetic biological fluids (saliva, urine, and vitreous humour) yielded recoveries close to 100% with negligible matrix effects. The bulk incorporation of BG into the ink formulation provided reproducible electrode fabrication, supporting the use of bulk-modified SPE for on-site forensic screening, while allowing both analytical responses to be generated from the same sensing interface. The bulk-modified SPE integrates visual screening and electrochemical confirmation and quantification within a single portable sensing platform, enabling complementary analytical information to be obtained from the same sensing interface without requiring addition reagents.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.