RSC advances

Paper-based hydrogel biosensor integrated with carbon dots for portable detection of microRNA-21 using a fluorescent diffusion area as the readout.

Xiaoqing Chen, Wansheng Lin, Qiang Lin, Bin Qiu, Zhixin Li, Chunjian Ye, Jianqing Liu

Department of Nephrology, Fuzong Clinical Medical College of Fujian Medical University Fuzhou Fujian 350025 China 534925185@qq.com. · Department of Cardiology, Fuzong Teaching Hospital of Fujian University of Traditional Chinese Medicine 156 Xierhuan Northern Road Fuzhou Fujian 350025 China.

Key points

From the paper's conclusion
  • The miR-21-responsive water-absorption hydrogel loaded with carbon dots (CDs) synthesized by sodium hyaluronate and the complementary capturing DNA (cDNA) of miR-21 is constructed for fluorescent area generation on filter paper.

Abstract

The early detection of diabetic kidney disease (DKD) is limited by the lack of suitable biomarkers. Circulating microRNA-21 (miR-21), a key pathogenic regulator correlated with early injury, presents a promising solution for point-of-care testing. In this work, a paper-based hydrogel biosensor that relies on fluorescent areas as the readout has been constructed for the convenient and ultrasensitive quantification of miR-21. The miR-21-responsive water-absorption hydrogel loaded with carbon dots (CDs) synthesized by sodium hyaluronate and the complementary capturing DNA (cDNA) of miR-21 is constructed for fluorescent area generation on filter paper. The miR-21 can be captured by the cDNA in a hydrogel, and the cDNA in the miR-21-cDNA duplex is cleaved by a duplex-specific nuclease. Then, the released miR-21 can be captured by other cDNA, thereby achieving signal amplification. The hydrogel was degraded via the cleavage of cDNA, and the CD-containing water was then released from the hydrogel fragments and spread uniformly on filter paper, thereby forming a distinct fluorescent region. The fluorescent area can be imaged by a smartphone and quantified via the pixel ratio of the fluorescent region to the whole paper area, which is then converted to the actual physical area. The developed biosensor demonstrates a linear response to miR-21 ranging from 1 pM to 10 nM with a detection limit of 131.25 fM. The portable sensing method exhibits portability, convenience, and high selectivity and has been applied to detect miR-21 in serum samples with satisfactory results, showing promising potential for early DKD screening.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers