Department of Medical Devices Engineering, Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvananthapuram, Kerala 695011, India. manojgopi@sctimst.ac.in. · Central Analytical Facility, Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvananthapuram, India.

Key points

From the paper's conclusion
  • Here, we report a label-free, reagent minimised electrochemical immunosensor fabricated by depositing carboxyl functionalised graphene oxide (GO) on fluorine-doped tin oxide (FTO) glass electrodes, followed by covalent antibody immobilization via EDC-NHS-PEG chemistry.

Abstract

Heart failure (HF) remains a leading cause of global morbidity and mortality, with timely and accurate biomarker quantification being critical for diagnosis, risk stratification, and therapeutic monitoring. N-terminal pro-B-type natriuretic peptide (NT-proBNP) is the most clinically reliable biomarker for HF; however, current detection platforms, ranging from chemiluminescent immunoassays to lateral flow devices, are limited by high infrastructure requirements, enzymatic labels, complex fabrication workflows, or insufficient sensitivity at physiologically critical concentrations. Here, we report a label-free, reagent minimised electrochemical immunosensor fabricated by depositing carboxyl functionalised graphene oxide (GO) on fluorine-doped tin oxide (FTO) glass electrodes, followed by covalent antibody immobilization via EDC-NHS-PEG chemistry. The platform exploits the oxygen containing functional groups on GO to enable stable antibody immobilisation without additional signal amplifying nanomaterials. The immunosensor was characterised using UV-vis spectroscopy, Fourier transform infrared (FTIR) spectroscopy, zeta potential analysis, scanning electron microscopy (SEM), atomic force microscopy (AFM), and confocal Raman chemical mapping, collectively confirming the formation of a uniform GO film and successful antibody immobilisation. Electrochemical characterization by cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) demonstrated that a GO coating concentration of 3 mg mL-1 optimally enhanced charge transfer, increasing the electrochemically active surface area by ∼15% relative to bare FTO. Antigen sensing was performed by monitoring the concentration-dependent increase in charge-transfer resistance (RCT) via EIS across a physiologically and clinically relevant NT-proBNP concentration range of 100 pg mL-1 to 500 ng mL-1, yielding a linear regression of Y = 0.0038x + 676.96 (R2 = 0.997) and a limit of quantification (LOQ) of 100 pg mL-1 directly spanning the diagnostic threshold for HF. The key novelty of this work lies in the synergistic use of the GO-FTO interface, which eliminates the need for additional nanoparticle enhancement, enzymatic signal amplification, or complex multi-layer fabrication, while still delivering clinically actionable sensitivity. The platform is scalable, cost-effective, and compatible with portable electrochemical instrumentation, establishing a strong foundation for point-of-care integration. These findings provide a promising foundation for future optimisation of the GO-FTO immunosensor, including clinical sample validation, device integration, and translation towards point-of-care heart failure diagnostics.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Design and analytical validation of a graphene oxide-modified FTO immunosensor for NT-proBNP detection — Science Skim