Analytical methods : advancing methods and applications

Integrated analytical quality by design enabled stability-indicating LC-MS method for cabazitaxel: forced degradation profiling, in silico pharmacokinetic prediction, and multi-metric greenness assessment.

Venkatesh Bukkapatnam, Siva Nageswara Rao Gajula, Mathrusri Annapurna Mukthinuthalapati, Raja Sundararajan

Department of Pharmaceutical Analysis, GITAM School of Pharmacy, GITAM - Deemed to be University, Rushikonda, Visakhapatnam 530 045, Andhra Pradesh, India. sraja61@gmail.com.

Computational

Key points

From the paper's conclusion
  • Overall, the developed integrated approach provides a reliable platform for CBZ quality control, stability assessment, and degradation profiling.

Abstract

Cabazitaxel (CBZ) is a second-generation semi-synthetic taxane used for metastatic castration-resistant prostate cancer therapy. This study aims to introduce a novel, streamlined analytical workflow for CBZ that bridges a critical gap in current literature by uniquely integrating Analytical Quality by Design (AQbD) chromatographic optimization, stability-indicating forced degradation, LC-MS degradation-product profiling, in silico absorption, distribution, metabolism, excretion, and toxicity (ADMET) evaluation, and a multi-metric greenness assessment into a single, comprehensive pipeline. Separation was achieved on a Phenomenex Luna C8 column using 0.1% acetic acid in water : acetonitrile (38 : 62, v/v) at 0.8 mL min-1 with UV detection at 229 nm. A Box-Behnken design was used to optimize critical method parameters, under which CBZ was eluted at approximately 12 min. Forced degradation included acidic hydrolysis (0.1 N HCl, 80 °C, 30 min), alkaline hydrolysis (0.01 N NaOH, 50 °C, 30 min), oxidation (3% H2O2, 80 °C, 30 min), dry heat (80 °C, 5 days), and ICH Q1B photolysis. The greatest degradation occurred under alkaline (33.33%), photolytic (32.55%), and oxidative (29.92%) stress. Four major degradation products (DPs) were tentatively assigned based on LC-MS data at m/z 735.74, 600.85, 572.77, and 775.73. In silico ADMET profiling for CBZ and its DPs showed that DP2 crosses the blood-brain barrier. The method was linear over 0.06-270 µg mL-1 (r2 = 0.9998), with LOD and LOQ values of 0.02 and 0.06 µg mL-1, respectively; recovery was 98.84-99.69%, and precision was <2% RSD. The method was applied to marketed injections, giving assay recoveries of 98.47-98.89%. Greenness/applicability assessment yielded an Eco-Scale score of 90, AGREE (0.61), BAGI (80.0), RAPI (80.0) and AGSA (84.72); GAPI identified workflow-specific environmental trade-offs, while Mosaic-WAC gave an overall score of 82.3. These complementary metrics support favorable overall analytical performance. Overall, the developed integrated approach provides a reliable platform for CBZ quality control, stability assessment, and degradation profiling.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Integrated analytical quality by design enabled stability-indicating LC-MS method for cabazitaxel: forced degradation profiling, in silico pharmacokinetic prediction, and multi-metric greenness assessment — Science Skim