Analytical methods : advancing methods and applications

LC-MS and UV spectroscopy comparative analysis of bisphenol A in plastic-packaged milk.

Kumarasamy Jayakumar, Habibulla Imran, Sooman Lim, Subramaniyan Ramasundaram, Tae Hwan Oh

Department of Biomaterials, Saveetha Dental College and Hospitals, SIMATS, Saveetha University, Chennai-600077, India. jkvare@gmail.com. · Department of Flexible and Printable Electronics, LANL-JBNU Engineering Institute-Korea, Jeonbuk National University, Jeonju 54896, Republic of Korea.

Key points

From the paper's conclusion
  • These findings suggest that UVS can provide higher detection sensitivity than LC-MS under the present experimental conditions, supporting BPA monitoring and food-safety assessment.

Abstract

Bisphenol A (BPA), a known endocrine disruptor used in food-grade plastics, was effectively eliminated from milk samples using a simple, sensitive, and efficient dispersive liquid-liquid extraction (LLE) method. Liquid chromatography-mass spectrometry (LC-MS) and ultraviolet spectroscopy (UVS) were used to evaluate BPA detection sensitivity and determine BPA concentrations in cow and buffalo milk samples. With LC-MS, reliable quantification, high precision, and nanogram-level detection were achieved, with a retention time of 3.58 min. In the UVS studies, BPA was measured over a concentration range of 0.43 nM to 50 µM, with a limit of detection (LOD) of 0.87 nM. LC-MS was used to detect BPA in the linear range of 1 nM to 1 µM, and the LOD and LOQ were 2.2 nM (S/N = 3) and 6.67 nM, respectively. This method exhibited good linearity (R2 = 0.998), satisfactory recoveries (91-102%), acceptable matrix effects (83.48-102.81%), and high precision and reproducibility with intra-day and inter-day RSDs of 0.84-2.31% and 1.09-3.18%, respectively. BPA concentrations in cow milk in India (0.5 nM) were significantly lower than levels reported in China, Japan, Korea, Europe, and the U.S. Continued BPA monitoring and the adoption of BPA-free packaging are essential. These findings suggest that UVS can provide higher detection sensitivity than LC-MS under the present experimental conditions, supporting BPA monitoring and food-safety assessment.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

LC-MS and UV spectroscopy comparative analysis of bisphenol A in plastic-packaged milk — Science Skim