Thickness-matched FT-IR benchmarking reveals spectral bias in AFM-IR measurements of supported silica films.

Tatsuhiro Nagasaka, Keiko Inoue, Aki Suzuki, Yusaku Tanahashi, Hirofumi Seki

Toray Research Center, Inc., 2-11, Sonoyama 3-chome, Otsu, Shiga 520-8567, Japan. tatsuhiro.nagasaka.w5@trc.toray.

Computational

Key points

From the paper's conclusion
  • Thickness-matched far-field benchmarking combined with model-independent line-shape analysis provides a practical framework for evaluating the measurement response embedded in AFM-IR peak changes before attributing them to nanoscale chemical heterogeneity.

Abstract

Atomic force microscopy-infrared (AFM-IR) spectroscopy enables nanoscale chemical characterization, but apparent peak shifts in supported thin films can combine genuine film-structure evolution with thickness-dependent measurement response. Here, replicated AFM-IR spectra of TEOS-derived silica films with nominal thicknesses of 10 nm, 50 nm, 100 nm and 200 nm were benchmarked against thickness-matched transmission Fourier-transform infrared (FT-IR) spectra. Film thicknesses were determined by X-ray reflectivity. The principal Si-O-Si band shifted to lower wavenumber with increasing thickness in both methods, but the total shift was approximately 23 cm-1 in AFM-IR and only 6 cm-1 in FT-IR. The AFM-IR thickness dependence therefore substantially exceeded the trend shared by the two methods, indicating additional technique-specific spectral weighting. A high-wavenumber area fraction defined without component fitting showed the largest excess in AFM-IR relative to FT-IR for the 10 nm film, which also exhibited the largest between-method peak-position difference. The AFM-IR/FT-IR area-fraction ratio decreased from 2.07 at 10 nm to 1.06-1.08 at 100 nm to 200 nm, and this thickness trend was retained when the baseline and integration windows were varied. FT-IR/XRR-derived SiOH number-density trends were also consistent with an AFM-IR-resolved SiOH-related O-H feature near 3620 cm-1 in the 50 nm, 100 nm and 200 nm films. Thickness-matched far-field benchmarking combined with model-independent line-shape analysis provides a practical framework for evaluating the measurement response embedded in AFM-IR peak changes before attributing them to nanoscale chemical heterogeneity.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Thickness-matched FT-IR benchmarking reveals spectral bias in AFM-IR measurements of supported silica films — Science Skim