From the paper's conclusion- Here, we present the design, experimental testing, and proof-of-concept validation of a passive acoustic Helmholtz resonator cavity for heartbeat sensing in the [50-120] Hz frequency range and determination of inter-beat interval (IBI) in a non-invasive manner using sound.
Abstract
Physiological markers are a primary source of prognostic health monitoring systems. Specifically, with heartbeat sensing, many ailments may be identified in their nascent stage. In clinical settings, the human heartbeat is measured by way of an electrocardiogram (ECG). Common metrics for heartbeat analysis include the heart rate (HR), or average beat-to-beat value over time, and heart rate variability (HRV), or variation of beat-to-beat intervals over time. In day-to-day cardiac monitoring, contact-based sensing techniques are realized using various wearable technologies. Here, we present the design, experimental testing, and proof-of-concept validation of a passive acoustic Helmholtz resonator cavity for heartbeat sensing in the [50-120] Hz frequency range and determination of inter-beat interval (IBI) in a non-invasive manner using sound. Integral to this sensing process, a unique neural network-based machine learning algorithm is exploited to reconstruct the ECG waveform from the resonator-enhanced acoustic signal. The system achieved a HR estimation error of 4.23 ± 4.50 BPM and an R-peak timing mean absolute error (MAE) of 36.50 ± 53.17 ms relative to a simultaneously acquired reference ECG. The potential extension of the engineered sensor hardware plus software system for future non-invasive cardiac biomonitoring of seated users in a dynamic vehicle environment is discussed. Paul Schmalenberg and colleagues present the design, testing, and validation of an acoustic Helmholtz resonator cavity for heartbeat sensing. A neural network-based algorithm accurately reconstructs the electrocardiogram waveform from the measured acoustic signal, which enables cardiac biomonitoring of seated users in dynamic vehicle environments.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- H ygieia : A Wearable Metamaterial Patch for Calibration-free Carotid Hemodynamics MonitoringProceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies · 2026
- Broadband wearable acoustic sensing for continuous cardiopulmonary monitoring.Microsystems & nanoengineering · 2026
- Heart Rate Variability Estimation from 3D Seismocardiography: A Deep Learning Approach with a Large-Scale Dataset.Physiological measurement · 2026
- Multiscale temporal deep learning for ECG waveform estimation from contactless radar signals.Biomedical physics & engineering express · 2026
- Millimeter-wave radar for non-contact heart rate monitoring: Composite filtering and multidimensional evaluation.Technology and health care : official journal of the European Society for Engineering and Medicine · 2026
- Wearable and Invisible ECG Quality and Usability Assessment in Cockpit Monitoring.Biosensors · 2026