A point-of-care lateral flow assay for semiquantitative detection of vitamin A deficiency for low-resource settings

Lavita Pinto, Kavitha S. Shettigar, Harshitha Mahesh, Srinivas Hebbar, Gayathri M Rao, Dinesh Upadhya

Manipal Academy of Higher Education

Key points

From the paper's conclusion
  • The presented semiquantitative LFA demonstrates promising results for a rapid, simple, economical, and point-of-care screening for VAD, suitable for limited-resource settings.

Abstract

Abstract Vitamin A is essential for vision, neurodevelopment, immunity and other functions. Vitamin A deficiency (VAD) remains prevalent in low-resource settings and vulnerable populations, particularly during pregnancy and lactation and in children, due to the unavailability of advanced detection methods. An inexpensive point-of-care semi-quantitative method could support early detection of VAD in such conditions. This study used a gold nanoparticle (AuNP)- based lateral flow assay (LFA) to detect VAD, with retinol-binding protein (RBP) as a Marker. The Nanoparticle Tracking Analysis and NanoDrop were used to characterize AuNPs conjugated with anti-RBP antibody and to determine the optimal conjugation pH, respectively. Antibody concentrations were optimized on nitrocellulose membranes to achieve reproducible, semi-quantitative visual signals. The values obtained from the LFA semiquantitative kit were confirmed and validated by ELISA, and the developed color chart was evaluated for sensitivity, specificity, and accuracy. Optimal conditions for antibody conjugation were achieved at pH 8.5, and 0.5–1 µg of the capture antibody resulted in the formation of the test line. The LFA demonstrated good agreement with ELISA in distinguishing low RBP from normal RBP levels, enabling classification into normal (> 19 µg/mL), subclinical (12–19 µg/mL), and clinical deficiency (< 12 µg/mL) categories, with 93.70% accuracy, 93.95% sensitivity, and 92.73% specificity. The presented semiquantitative LFA demonstrates promising results for a rapid, simple, economical, and point-of-care screening for VAD, suitable for limited-resource settings. While clinical verification is needed to validate the clinical cutoffs and the test’s feasibility for Point-of-care screening, this kit provides an opportunity to develop several semiquantitative kits for other biomarker proteins.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A point-of-care lateral flow assay for semiquantitative detection of vitamin A deficiency for low-resource settings — Science Skim