Investigation of the dosimetric characteristics of real-time silicone/Bi2O3 nanocomposites for the diagnostic level X-rays

Farideh Keyvani, Leila Gholamzadeh, Shahryar Malekie, Sedigheh Kashian

Key points

From the paper's conclusion
  • This study presents the development of high-sensitivity real-time X-ray dosimeters based on room-temperature-vulcanized (RTV-2) silicone nanocomposites incorporating 0–60 wt% Bi 2 O 3 nanoparticles, fabricated via solution casting.

Abstract

This study presents the development of high-sensitivity real-time X-ray dosimeters based on room-temperature-vulcanized (RTV-2) silicone nanocomposites incorporating 0–60 wt% Bi 2 O 3 nanoparticles, fabricated via solution casting. Structural and morphological characterization using X-ray diffraction (XRD), scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDX) with elemental mapping, and elemental analysis confirmed the uniform dispersion of monoclinic Bi 2 O 3 nanorods within the amorphous RTV-2 matrix. Complementary Fourier-transform infrared spectroscopy (FTIR), Raman spectroscopy, and thermogravimetric analysis (TGA) validated the chemical integrity, thermal stability, and mechanical robustness of the resulting nanocomposites. Dosimetric performance was investigated under diagnostic-range X-ray exposure using a COMET MXR 320/26 source (dose rates: 2.8–12 mGy/min). The I-V characteristics at 4.8 mGy/min exhibited a highly linear response over ± 400 V. At 400 V, the photocurrent increased from ~ 0.56 nA (pure RTV-2) to ~ 1.66 nA (60 wt% Bi 2 O 3 ), corresponding to a ~ 3-fold enhancement in photocurrent. The devices demonstrated good short-term repeatability (standard deviation of 2.3% over ten consecutive measurements) and a signal-to-noise ratio (SNR) of up to ~ 1900 at 400 V. Compared with reported values for commercial MOSFET dosimeters (~ 0.1–1 nA/mGy) and previously reported Bi 2 O 3 -polymer systems, the developed nanocomposites exhibit a higher competitive normalized sensitivity (~ 0.35 nA/(mGy/min), or 878 nC·mGy⁻ 1 ·cm⁻ 3 on a volumetric basis), demonstrating their suitability as lead-free alternatives for low-dose, real-time diagnostic X-ray dosimetry; further evaluation of long-term stability and mechanical durability is needed to establish clinical applicability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers