Scientific Reports

Optimizing fixation and staining to reduce cell detachment in wet-fixed direct smears, a quantitative evidence-based quality improvement study

Yinying Zou, Xin Wang, Yan Chen, Zihan Wu

Key points

From the paper's conclusion
  • An evidence-based strategy for laboratory quality improvement—extending fixation toward 3 h, including an intermediate alcohol step (preferably 50% ethanol) at each transition between water and 95% ethanol, and using ordinary slides—substantially reduces detachment.

Abstract

Wet-fixed cytologic smears are inherently prone to cell detachment, which causes cross-contamination and reduces smear cellularity, potentially leading to false-positive diagnoses and diagnostic uncertainty. This study evaluated the effects of the intermediate alcohol step, fixation duration, and slide type on detachment to establish an evidence-based optimized protocol. We evaluated: (i) the necessity and optimal concentration of the intermediate alcohol step; (ii) fixation duration from 30 min to 24 h; and (iii) slide type (ordinary vs. positively charged). Detachment was scored semi-quantitatively using a streak-wiping method. Wilcoxon and Friedman tests were used for paired and block comparisons, respectively. The intermediate alcohol step significantly reduced detachment ( p < 0.05), with 50% ethanol superior to 75% ( p = 0.012). Prolonged fixation progressively decreased detachment ( p < 0.001), with no further statistically significant reduction beyond 3 h. Under optimized conditions, ordinary and positively charged slides performed comparably ( p = 0.811). An evidence-based strategy for laboratory quality improvement—extending fixation toward 3 h, including an intermediate alcohol step (preferably 50% ethanol) at each transition between water and 95% ethanol, and using ordinary slides—substantially reduces detachment. By preserving cellularity and helping to reduce the risk of cross-contamination, this protocol offers an implementable quality improvement tool for cytology laboratories, potentially enhancing diagnostic reliability and patient safety in wet-fixed cytologic preparations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Optimizing fixation and staining to reduce cell detachment in wet-fixed direct smears, a quantitative evidence-based quality improvement study — Science Skim