Biomechanical effects of length and angle of the subcutaneous tunnel on epidural catheter stability in a porcine model.
Jan Pennig, Sarah N Schmidt, Axel Schmutz, Johannes Hell
Department of Anesthesiology and Critical Care, Medical Center, University of Freiburg, Hugstetter Str. 55, 79106, Freiburg, Germany. · Department of Anesthesiology and Critical Care, Medical Center, University of Freiburg, Hugstetter Str. 55, 79106, Freiburg, Germany. johannes.hell@uniklinik-freiburg.de.
Key points
From the paper's conclusion- In conclusion, subcutaneous tunneling of epidural catheters provides limited mechanical benefit on catheter stability under standardized conditions in a porcine model whereas adhesive fixation generates substantially greater resistance to immediate catheter dislocation.
Abstract
Epidural catheter dislocation can lead to severe complications, especially in anticoagulated patients. Fixation techniques vary, and the role of subcutaneous tunneling and the biomechanical effects of length or angle towards dislocation forces of the tunnel remain unknown. This in vitro study evaluated the contribution of length and angle of subcutaneous tunneling (2-5 cm at six various angles relative to dislocation force) compared to different adhesive fixation methods (transparent film dressing, adhesive strips 10-20 cm) on epidural catheter stability. Dislocation forces were measured in a porcine skin model using a motorized traction system and a force transducer. The mean dislodgement force for tunneled catheters was 0.86 ± 0.3 N. This was significantly lower than the forces measured using transparent film dressing (9.89 ± 2.65 N; p < 0.001) or adhesive strips (10 cm: 8.03 ± 2.3 N; 20 cm: 10.92 ± 0.84 N; all p < 0.001). Tunnel length and angle had no significant impact on catheter stability whereas increasing the length of the adhesive strip significantly increased the dislodgement force (p < 0.001). In conclusion, subcutaneous tunneling of epidural catheters provides limited mechanical benefit on catheter stability under standardized conditions in a porcine model whereas adhesive fixation generates substantially greater resistance to immediate catheter dislocation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Two-point Loop Suture Fixation May Reduce Catheter Displacement in Cisternal and External Ventricular Drains.Neurologia medico-chirurgica · 2026
- Mechanical stability of endotracheal tube fixation: a performance comparison between two tube holders and a conventional method.BMC anesthesiology · 2026
- Epidural Fibrosis Raises Thresholds but Preserves Recruitment in Conventional Spinal Cord Stimulation: Lateral Steering Is Fragile.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026
- Dual-step burr hole with periosteal anchoring technique: a robust method for ventricular catheter fixation - how I do it.Acta neurochirurgica · 2026
- A Porcine Abdominal Model for Quantitative Assessment of Incision Tension After Sequential Subcutaneous Dissection.Journal of visualized experiments : JoVE · 2026
- Management of Retained Epidural Catheter Fragment in a Polytrauma Patient: A Case Report.Journal of orthopaedic case reports · 2026