Towards prospective identification of optimal scan parameters for CT-guided interventions: a phantom study.
Constantin Schareck, Neele Blumhoff, Malte M Sieren, Roman Kloeckner, Franz Wegner
Institute for Interventional Radiology, University of Lübeck, Ratzeburger Allee 160, 23538, Lübeck, Germany. constantin.schareck@uksh.de. · Institute for Interventional Radiology, University of Lübeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.
Key points
From the paper's conclusion- In conclusion, for optimal image quality in CT-guided interventions, scan protocols must adapt tube voltage and tube load dynamically to the target tissue and patient size.
Abstract
CT-guided interventions require balance between radiation dose and image quality. This study evaluates the impact of tube voltage and tube load on the contrast-to-noise ratio (CNR) across different patient sizes and target tissues. A modular multi-energy phantom with effective diameters of 20 cm and 35 cm was used. Bone, iodine-blood mixture, and pure blood targets were scanned axially at matched dose levels. Images were equally reconstructed. For analysis, CNR was the primary metric. Tube voltages of 80, 100, 120 and 140 kVp were applied with corresponding mAs adjustments to achieve dose levels between 1 and 10 mGy SSDE (Size Specific Dose Estimate). In the small phantom, low kVp settings yielded the highest CNR for high-contrast targets like iodine. For low-contrast imaging, higher kVp settings were more beneficial. In the large phantom, higher kVp consistently improved CNR across all materials. E.g. raising kVp from 80 to 120 raised the CNR from 6.5 to 8.6 for the iodine-blood setup. In conclusion, for optimal image quality in CT-guided interventions, scan protocols must adapt tube voltage and tube load dynamically to the target tissue and patient size. Finding the sweet spot between contrast loss and noise reduction for increasing tube voltages is key for intervention planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Radiation and Contrast Medium Dosage Optimization of Coronary CT Angiography While Combining Personalized Patient Protocol Technology and Automated Tube Voltage Selection Technology.Diagnostics (Basel, Switzerland) · 2026
- Image quality and radiation dose optimization in CT pulmonary angiography using 120, 100, and 80 kV protocols for pulmonary thromboembolism diagnosis.Quantitative imaging in medicine and surgery · 2026
- Influence of X-ray Tube Voltage on Computed Tomography Values and Contrast-Noise Ratios in Virtual Monoenergetic Images in routine clinical Photon Counting Detector Computed Tomography: Does kV Still Matter?Investigative radiology · 2026
- Quantitative comparison of 68-keV monoenergetic, routine-dose, and low-dose 120-kVp CT for calcium-mimicking inserts in thoracic phantoms: effects of phantom size and reconstruction method.European radiology experimental · 2026
- Assessment of image quality and radiation dose in low-dose, low-contrast computed tomography coronary angiogram: a systematic review and meta-analysis.Clinical radiology · 2026
- Size-specific weighted detectability index for computed tomography image characterization.Medical physics · 2026