Validation of a novel 3D-printed anthropomorphic pediatric abdomen phantom using photon-counting CT
Thilo Schikorra, Matthias May, Thomas Flohr, Thomas Allmendinger, Ferdinand Knieling, Ido Bitan, R. C. Reina, Michael Uder,
Friedrich-Alexander-Universität Erlangen-Nürnberg · Siemens Healthcare (United States)
Key points
From the paper's conclusion- This study compares objective and subjective image quality (IQ) between a pediatric patient's CT scan and a subsequently 3D-printed anthropomorphic abdomen CT phantom on a photon-counting CT (PCCT).
Abstract
This study compares objective and subjective image quality (IQ) between a pediatric patient's CT scan and a subsequently 3D-printed anthropomorphic abdomen CT phantom on a photon-counting CT (PCCT). We used a dedicated 3D-printed CT phantom based on DICOM data obtained from a PCCT scan of a 5-year-old boy (70 kV, 458 mAs). We performed two scans on the phantom (1: 70 kV, 461 mAs and 2: 90 kV, 148 mAs), reconstructing them with three kernels. We measured the objective IQ based on mean CT values and image noise within eight target structures. Three radiology experts assessed the subjective IQ on a 5-point Likert scale. The CT values of organs and vessels in phantom scan 1 (Br40) were equivalent to those in the patient (p = 0.019). The phantom demonstrated higher mean CT values for fat and lower CT values for bone. The image noise was equivalent between both scans (p = 0.027). The phantom showed high subjective IQ but a significantly lower rating than the patient (median 4 vs. 5, p < 0.001). In conclusion, this study demonstrates comparable IQ between the phantom and the patient scan at comparable dose parameters. In the future, it may be possible to anticipate radiation exposure in dose reduction studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Can CT Phantoms Produce Realistic CT Images of Human Anatomy? Physics-Based Evaluation of Multimaterial 3D-Printed CT Phantoms From CT Data.Investigative radiology · 2026
- Photon-counting CT in pediatric patients: A multi-institutional evaluation of protocols and radiation exposure in 4,772 patientsDiagnostic and Interventional Imaging · 2025
- Optimization of the reconstruction kernel for temporal bone imaging using photon-counting detector CT: A combined physical and visual evaluationRadiography · 2025
- Photon-Counting CT Enables Higher Image Quality at a Lower Radiation Dose Compared with a 256-Slice Energy-Integrating Detector CT in Pediatric Patients with Congenital Heart DiseaseDiagnostics · 2026
- 3D Fabrication and Multiscanner CT Validation of a Novel Patient-Specific Anthropomorphic Pediatric Head Phantom.Academic radiology · 2026
- Development and validation of a novel three-dimensional patient-specific pediatric brain phantom for computed tomography neuroimaging research.Pediatric radiology · 2026