Scientific Data

Brain 18F-FDG template using images from a new brain-dedicated positron emission tomography scanner

Pedro Nespral, Gonzalo Cuesta-Domingo, María Fernanda Ollarves-Carrero, Blanca García-Raldúa, Mirjam Brackhan, Elena Marcos-Macías, Miguel Ángel Pozo, José Luis Carreras-Delgado,

Key points

From the paper's conclusion
  • We obtained an 18 F-FDG-PET brain template with 1-mm 3 voxels and a 268 x 268 x 152 matrix to use as a reference for PET image normalization.

Abstract

18 F-fluorodeoxyglucose positron emission tomography ( 18 F-FDG-PET) may be a useful tool as a biomarker of neurodegeneration. However, commercial brain templates may differ substantially from the PET images generated in new scanners in terms of resolution. We aimed to build an 18 F-FDG-PET image template using images acquired in a brain-dedicated PET to enable normalization optimization. A total of 167 patients underwent an 18 F-FDG-PET study on a brain-dedicated PET system. The images were automatically co-registered in Pmod 4.1, followed by iterative deformable registration in ANTs (Advanced Normalization Tools) to build the template. We obtained an 18 F-FDG-PET brain template with 1-mm 3 voxels and a 268 x 268 x 152 matrix to use as a reference for PET image normalization. In addition, we generated a map with 66 regions of interest (ROIs) manually adapted from the Automated Anatomical Labeling (AAL) ROI map for quantification of the normalized images and a mask to exclude extra-parenchymal regions. The development of a brain template with images acquired on the dedicated brain PET is of great utility for application in new scanners with improved resolution.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Brain 18F-FDG template using images from a new brain-dedicated positron emission tomography scanner — Science Skim