A soft bioelectronic system with a responsive hydrogel neural interface for direct force quantification and real-time intraoperative craniocerebral injury monitoring.
Bo Pang, Ganguang Yang, Qihong Cheng, Sen Zhou, Zhixin Wang, Wang Xuan, Yuqi Qiu, Hangyu Gong,
Flexible Electronics Research Center, State Key Laboratory of Intelligent Manufacturing Equipment and Technology, School of Mechanical Science and Engineering, Huazhong University of Science and Technology, Wuhan, China. · Department of Neurosurgery, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
Key points
From the paper's conclusion- Here we report a soft neural interface bioelectronic system (SNIBS) for real-time monitoring of craniocerebral injury throughout neurosurgery.
Abstract
In craniotomies, the surgical manipulation of instruments inherently involves applying mechanical forces to brain tissues, potentially causing brain damage. Currently, there is no direct method to quantitatively monitor craniocerebral injuries. Here we report a soft neural interface bioelectronic system (SNIBS) for real-time monitoring of craniocerebral injury throughout neurosurgery. On the basis of the thermally triggered swelling/deswelling of hydrogel microspheres, the hydrogel interface achieves seamless integration with the brain cortex for recording high-fidelity signals during surgery while ensuring benign detachment to prevent brain damage after surgery. The SNIBS enables the monitoring of contact pressure, somatosensory evoked potentials, electrocorticography and temperature, providing a direct method of assessing neural status. The correlation between somatosensory evoked potential and cortical neural pathway damage caused by pressure is revealed by SNIBS. Evans blue staining and magnetic resonance imaging confirm that contact pressure higher than 80 kPa causes irreversible craniocerebral injury. Validated in rabbit brain tumour excision models, the SNIBS enables surgeons to quantify forces to avoid craniocerebral injury. The presented technology holds strong potential for establishing safety evaluation standards for neurosurgery.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A soft electrode array with reconfigurable hydrogel interfaces for high-fidelity neurophysiological monitoring during craniotomyProceedings of the National Academy of Sciences · 2026
- Implanted intracranial pressure sensors: mechanisms, fabrication, and system integration.Nanoscale horizons · 2026
- Dynamic Conformal Neural Electroactive Biointerfacing Sustains Bioelectronic Coupling for Spinal Cord Repair.ACS nano · 2026
- A Neurocompliant Ionic Conductive Hydrogel Orchestrates Glial and Vascular Remodeling to Mitigate Traumatic Brain InjurybioRxiv · 2026
- Injectable Bioadhesive Ultrasonic Sensor for Continuous and Wireless On-Demand Pressure Monitoring in Acute Compartment Syndrome.Advanced materials (Deerfield Beach, Fla.) · 2026
- Responsive Microgel-Reinforced Multiple Dynamic Cross-Linked Hydrogels with High Toughness and Low Hysteresis for Bioelectronic SensorBiomacromolecules · 2025