Automated retinal layer thickness segmentation in neovascular age-related macular degeneration: a comparison of two algorithms.
Giacomo Boscia, Jonathan D Oakley, Tommaso Tibaldi, Daniel B Russakoff, Claudio Foti, Lorena Ulla, Paolo Caselgrandi, Francesco Gelormini,
Department of Surgical Sciences, University of Turin, Turin, Italy. · Voxeleron Inc, Austin, USA.
Key points
From the paper's conclusion- While correlations between the two segmentation algorithms were moderate to strong for several metrics, clinically meaningful mean differences were observed in specific compartments, most notably for central neurosensory and total retinal thickness. These findings indicate that the two algorithms should not be used interchangeably.
Abstract
PURPOSE: To evaluate the consistency of retinal measurements provided by two automated segmentation algorithms in eyes with neovascular age-related macular degeneration (AMD). DESIGN: Retrospective cohort study. PARTICIPANTS: Fifty patients with neovascular AMD. METHODS: Retinal thickness was measured using a Heidelberg Spectralis HRA + OCT spectral-domain device. Measurements were automatically averaged across ETDRS grid subfields centered on the fovea using both the Spectralis built-in software (version 1.10.4.0) and the Orion automated OCT segmentation software (Voxeleron, version 3.0.0). MAIN OUTCOME MEASURES: Differences in quantitative retinal thickness metrics between the two algorithms. RESULTS: For inner retinal thickness, the mean difference was 8.97 μm in the central fovea (R² = 0.32), with larger variability in the inner (22.09-28.73 μm; R² = 0.32-0.80) and outer circles (16.02-29.50 μm; R² = 0.45-0.41). Outer retinal thickness showed smaller differences, with a mean of 21.51 μm in the central fovea (R² = 0.36) and modest variation across subfields. The neurosensory retina demonstrated a mean difference of 0.13 μm centrally (R² = 0.66), increasing across ETDRS subfields, with an overall mean difference of 45.30μm (R² = 0.82). Total retinal thickness differed by 43.54 μm centrally (R² = 0.61), with smaller differences across the grid (mean 18.16 μm; R² = 0.60). CONCLUSIONS: While correlations between the two segmentation algorithms were moderate to strong for several metrics, clinically meaningful mean differences were observed in specific compartments, most notably for central neurosensory and total retinal thickness. These findings indicate that the two algorithms should not be used interchangeably.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Structure-Function Link of Ellipsoid Zone Integrity and Outer Retinal Features with Visual Acuity in Dry AMD.Diagnostics (Basel, Switzerland) · 2026
- Inter-individual biological variability of retinal layer thickness on optical coherence tomography scans.International ophthalmology · 2026
- Inner Retinal Reflectivity Loss and Concurrent Thickening Predict Short-Term Conversion to Neovascular AMDInvestigative Ophthalmology & Visual Science · 2026
- Longitudinal Changes in Inner Retinal Layers in Treatment-Naïve Neovascular AMD During Intravitreal Anti-VEGF Therapy.Ophthalmology. Retina · 2026
- Geometric and concavity characteristics as diagnostic markers for differentiating nAMD-related MNV from pachychoroid neovascularisation.Eye (London, England) · 2026
- Retinal fluid segmentation in real-world OCT imaging of neovascular AMD using hybrid deep learning and graph-based optimization.Biomedical optics express · 2026