Automated retinal layer thickness segmentation in neovascular age-related macular degeneration: a comparison of two algorithms.

Giacomo Boscia, Jonathan D Oakley, Tommaso Tibaldi, Daniel B Russakoff, Claudio Foti, Lorena Ulla, Paolo Caselgrandi, Francesco Gelormini,

Department of Surgical Sciences, University of Turin, Turin, Italy. · Voxeleron Inc, Austin, USA.

Observational

Key points

From the paper's conclusion
  • While correlations between the two segmentation algorithms were moderate to strong for several metrics, clinically meaningful mean differences were observed in specific compartments, most notably for central neurosensory and total retinal thickness. These findings indicate that the two algorithms should not be used interchangeably.

Abstract

PURPOSE: To evaluate the consistency of retinal measurements provided by two automated segmentation algorithms in eyes with neovascular age-related macular degeneration (AMD). DESIGN: Retrospective cohort study. PARTICIPANTS: Fifty patients with neovascular AMD. METHODS: Retinal thickness was measured using a Heidelberg Spectralis HRA + OCT spectral-domain device. Measurements were automatically averaged across ETDRS grid subfields centered on the fovea using both the Spectralis built-in software (version 1.10.4.0) and the Orion automated OCT segmentation software (Voxeleron, version 3.0.0). MAIN OUTCOME MEASURES: Differences in quantitative retinal thickness metrics between the two algorithms. RESULTS: For inner retinal thickness, the mean difference was 8.97 μm in the central fovea (R² = 0.32), with larger variability in the inner (22.09-28.73 μm; R² = 0.32-0.80) and outer circles (16.02-29.50 μm; R² = 0.45-0.41). Outer retinal thickness showed smaller differences, with a mean of 21.51 μm in the central fovea (R² = 0.36) and modest variation across subfields. The neurosensory retina demonstrated a mean difference of 0.13 μm centrally (R² = 0.66), increasing across ETDRS subfields, with an overall mean difference of 45.30μm (R² = 0.82). Total retinal thickness differed by 43.54 μm centrally (R² = 0.61), with smaller differences across the grid (mean 18.16 μm; R² = 0.60). CONCLUSIONS: While correlations between the two segmentation algorithms were moderate to strong for several metrics, clinically meaningful mean differences were observed in specific compartments, most notably for central neurosensory and total retinal thickness. These findings indicate that the two algorithms should not be used interchangeably.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Automated retinal layer thickness segmentation in neovascular age-related macular degeneration: a comparison of two algorithms — Science Skim