Agreement of wearable and standing bioimpedance devices with bioimpedance spectroscopy estimates of total body water in adults with Down syndrome.
Michael R Esco, Justo Perez, Elroy J Aguiar, Matthew A Mohammadnabi, Cara Chase, Kristen Heitman, Xueliang Pan, Stefan A Czerwinski,
Department of Kinesiology, College of Education, University of Alabama, Tuscaloosa, AL, USA. mresco@ua.edu. · School of Health and Rehabilitation Sciences, College of Medicine, The Ohio State University, Columbus, OH, USA.
Key points
From the paper's conclusion- These findings indicate that wearable bioimpedance technology demonstrated closer agreement with TBW estimates from BIS than standing BIA devices.
Abstract
BACKGROUND: Wearable bioimpedance technology offers a potentially accessible method for assessing total body water (TBW), yet its agreement with laboratory-based approaches in adults with Down syndrome remains unclear. The purpose of this study was to evaluate whether a wearable bioimpedance device demonstrates agreement with laboratory-based bioelectrical impedance spectroscopy (BIS) estimates of TBW comparable to standing bioimpedance devices in adults with Down syndrome. METHODS: Sixteen adults with Down syndrome (age: 28.3 ± 9.2 yr; height: 150.9 ± 7.8 cm; body mass: 77.4 ± 20.3 kg) completed TBW assessment using BIS (reference), a wearable bioimpedance device (BIAW), and two standing multi-frequency bioimpedance analyzers (BIAST1, BIAST2). RESULTS: Repeated measures ANOVA revealed significant differences among devices (F (3,42) = 17.52, p < 0.001). Both standing devices produced lower TBW values compared with BIS (p < 0.001), whereas BIAW did not differ from BIS (p = 0.106). Lin's concordance correlation coefficients demonstrated strong agreement between BIS and all field-based devices (CCC = 0.82-0.92), with the highest concordance observed for the wearable device. Bland-Altman analyses showed systematic underestimation for standing devices (-3.12 ± 4.65 L and -2.36 ± 4.64 L), while the wearable exhibited the smallest bias (-1.07 ± 4.64 L). Mean absolute percentage error ranged from 5.5% to 8.7%, with the wearable device demonstrating the lowest relative error. CONCLUSION: These findings indicate that wearable bioimpedance technology demonstrated closer agreement with TBW estimates from BIS than standing BIA devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement of the Watson and Hume-Weyers Equations with Total Body Water Derived from a Bioimpedance Scale Across Adiposity Levels in an Andean-Amazonian Population.Life (Basel, Switzerland) · 2026
- The Effect of Arm Position on Bioelectrical Impedance Measurements in Adults.Annals of biomedical engineering · 2026
- Examining within-session test-retest reliability of four bioimpedance devices among adults across a wide age range.Journal of electrical bioimpedance · 2026
- Measurement error in bioimpedance spectroscopy: The role of electrode type and analyzer.Journal of electrical bioimpedance · 2026
- Concordance of a Laboratory-Data Integrated Fluid Targeting Algorithm with Bioimpedance-Derived Volume Measures in Maintenance Hemodialysis.Kidney360 · 2026
- Wearables for health monitoring: body composition estimates of commercial smartwatch and clinical bioelectrical impedance deviceFrontiers in Sports and Active Living · 2025