Two-Dimensional Correlation Mass Spectrometry for Precursor/Fragment Ion Identification of Composite and Unresolved Spectral Peaks
Elly A. Breves, Garon H. Lowke, Jacob T. Shelley
Key points
From the paper's conclusion- These results show the utility of 2D-CMS for deconvolution of mass spectra acquired without separation or isolation of analytes or at low resolving powers.
Abstract
Two-dimensional correlation mass spectrometry (2D-CMS) is a cross-correlation-based technique that can be used to identify mass-spectral signals that arise from a common molecular origin. In this study, 2D-CMS is explored for mixtures of the blood-pressure medications atenolol and metoprolol with the specific aims of precursor/fragment-ion identification and deconvolution of unresolved spectral peaks. Atenolol and metoprolol are structurally similar and produce several identical or isobaric fragment ions. In addition, the first isotopic peak of the atenolol ion (i.e., [MH++1]) at m/z 268.1737 and the protonated metoprolol ion at m/z 268.1834 produce an isobaric overlap for the mass-spectral resolving power used in this study (m/Δm = 17,500 at m/z 200). These phenomena result in numerous peaks that originate from both analytes, complicating the interpretation of mass-spectral results for precursor/fragment identification and assignment of overlapped peaks. Semiquantitative information extracted from 2D-CMS results enabled the determination of the relative contribution of each analyte to the total signal of their shared fragments. In addition, narrow m/z regions within unresolved mass-spectral peaks could be attributed to their specific molecular origin. For a series of solutions with varying proportions of each analyte, metrics derived from 2D-CMS were found to scale with the relative concentration of each analyte. Calibration results obtained from 2D-CMS were comparable to those obtained from traditional peak-fitting methods but provided additional information that enabled peak identification. These results show the utility of 2D-CMS for deconvolution of mass spectra acquired without separation or isolation of analytes or at low resolving powers.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Integrated High‐Resolution Mass Spectrometry‐based Chemical Identification and Quantitative Analysis of Multiple Components With a Single Marker for Quality Evaluation of Dasiphora mandshurica (Maxim.) JuzSEPARATION SCIENCE PLUS · 2026
- Evaluating the Effects of Acquisition and Search Parameters on Repository-Scale MS/MS Matching.Journal of the American Society for Mass Spectrometry · 2026
- High‐Resolution Ion Mobility as an Alternative to Quadrupole‐Based Precursor Isolation for Reducing Chimeric Fragmentation Spectra in Bottom‐Up ProteomicsPROTEOMICS · 2025
- COSine: An Open, GUI-Based Tool for Robust Cosine-Similarity Scoring and Mirror Plotting of High-Resolution Mass Spectra MatchingAnalytica · 2026
- MS2-Based Evaluation of Precursor-Specific Fragmentation Conditions for Improving TMT Reporter Signal in MS3-Based QuantificationJournal of the American Society for Mass Spectrometry · 2026
- Evaluation of semi-quantification strategies for hindered amine light stabilizers by liquid chromatography-tandem mass spectrometry.Analytical methods : advancing methods and applications · 2026