Evaluating the Effect of Chromatographic Chelator Modifiers on Metabolite Levels in Untargeted Liquid Chromatography-Mass Spectrometry.
Daniel J Breiner, Vanessa V Phelan
Department of Pharmaceutical Sciences, Skaggs School of Pharmacy and Pharmaceutical Sciences, University of Colorado Anschutz, Aurora, Colorado80045, United States.
Key points
From the paper's conclusion- Collectively, our findings demonstrate that mobile phase chelators substantially influence untargeted metabolomic profiles beyond improving chromatography.
Abstract
Metal-sensitive metabolites frequently exhibit poor chromatographic performance during liquid chromatography-mass spectrometry (LC-MS) due to analyte interactions with stainless-steel components of the chromatographic system. Although mobile phase chelators are widely used to improve targeted analyses, their broader effects on untargeted metabolomics remain unclear. We systematically evaluated the impact of chromatographic chelator modifiers on untargeted LC-MS detection of Pseudomonas aeruginosa secondary metabolites using the metal-chelating quorum-sensing molecule Pseudomonas Quinolone Signal (PQS) as a model analyte. Analysis of publicly available metabolomics data sets revealed substantial variability in PQS chromatographic performance across laboratories. From this observation, we hypothesized that metal-mediated adsorption may contribute to inconsistent detection of PQS. We compared the effects of EDTA and citrate as mobile phase additives on P. aeruginosa metabolite detection across multiple culture media. Both chelators markedly improved PQS peak shape and enhanced detection of PQS and the siderophore pyochelin. However, high concentrations of EDTA caused ion suppression, altered adduct formation, and reduced detection of several metabolite classes, including phenazines and rhamnolipids, while citrate eliminated detection of phenazine-1-carboxylic acid (PCA). Reducing EDTA concentration to low micromolar levels largely preserved chromatographic improvements while minimizing adverse effects on metabolite coverage, although suppression of PCA persisted. These trends were consistent across chemically distinct growth media. Collectively, our findings demonstrate that mobile phase chelators substantially influence untargeted metabolomic profiles beyond improving chromatography. Careful selection of chelator identity and concentration is therefore essential to balance chromatographic performance with comprehensive metabolite detection in untargeted LC-MS analyses of metal-sensitive metabolites.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Matrix effects in untargeted LC-MS metabolomics: From creation to compensation with post-column infusion of standardsJournal of Chromatography A · 2025
- Aqueous phase leachates from pure and commercial microplastics differentially affect the physiology and metabolism of Chlorella sp. MM3.Environmental pollution (Barking, Essex : 1987) · 2026
- Sub-3 μm porous graphitic carbon for highly polar metabolomics: orthogonal selectivity and comparison with HILIC and reversed-phase chromatography.Analytical and bioanalytical chemistry · 2026
- Unexpected mitigation of microplastic-induced nitrification inhibition by sulfamethoxazole: The pivotal roles of polymer aging and microbial metabolic shifts.Journal of hazardous materials · 2026
- Multi-Layer Functional Metabolomics Reveals Chemical Interactions Between the Plant-Associated Bacterium Serratia marcescens and PhytophthorabioRxiv · 2026
- LC-MS-Based Chlorine-Signature Query Engine Strategy for Automated Discovery of Phosphate-Containing Metabolites.Analytical chemistry · 2026