From Benchtop to Body: Solution Resistance-Induced Distortion in Electrochemical, Aptamer-Based Sensors.
Jennifer M Gibson, Julian Gerson, Kon Son, Tod E Kippin, Kevin W Plaxco, Lior Sepunaru, Bob M Lansdorp
Interdisciplinary Program in Quantitative Biosciences, University of California, Santa Barbara, Santa Barbara, California93106, United States. · Nutromics Pty Ltd. , Brunswick, VIC3056, Australia.
Key points
From the paper's conclusion- Here, we show that uncompensated solution resistance of wire-based EAB sensors can increase from a few hundred ohms in vitro up to ∼1–2 kΩ in vivo.
Abstract
By providing continuous, in vivo molecular measurements, electrochemical aptamer-based (EAB) sensors have the potential to improve the fidelity with which pharmaceutical drugs are administered and physiological processes are monitored. The impact of solution resistance on EAB sensor response, however, has not been systematically investigated, despite its potential to introduce systematic errors in concentration measurements. Here, we show that uncompensated solution resistance of wire-based EAB sensors can increase from a few hundred ohms in vitro up to ∼1-2 kΩ in vivo. This elevated resistance increases the RC charging time constant of the electrode interface, allowing non-faradaic charging currents to persist into the sampling window used during square wave voltammetry. As a result, capacitive currents contribute to the measured signal, distorting sensor responses whenever the interrogation frequency falls within this RC-limited charging regime. Using three representative EAB sensors, we show that changes in uncompensated solution resistance between calibration and measurement conditions can, for some sensors, cause target concentrations to be systematically underestimated, with the magnitude of this error strongly dependent on interrogation frequency. We also introduce a simple strategy by which the calibration of in vivo EAB sensors can be improved by introducing a resistor in series with the working electrode during in vitro testing. Finally, we provide further strategies by which the errors introduced by uncompensated solution resistance can be reduced to negligible levels.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- On the Design, Fabrication, and Operation of Electrochemical Aptamer-Based SensorsACS Sensors · 2025
- General approach to achieving electrochemical aptamer-based sensor sensitivity of buffer in blood plasmanpj Biosensing · 2026
- The Long and Lucky Road to Continuous Molecular Monitoring: A Personal History of Electrochemical Aptamer-Based Sensors.Accounts of chemical research · 2026
- Self-Correction of pH-Induced Signal Variations in Methylene Blue-Labeled Aptamer Electrochemical Biosensors: Wearable Cortisol Detection in SweatACS Sensors · 2026
- Continuous Molecular Monitoring Using Electrochemical Aptamer-Based Sensors: Remaining Challenges for Long-Term In Vivo Deployment.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026
- Electrochemical Aptamer-Based Sensors Supporting Real-Time, Seconds-Resolved, Molecular Measurements in Awake, Behaving Rats.ACS sensors · 2026