ACS sensors

Capillary-Driven Surface-Enhanced Raman Scattering Hollow Microneedle Patch for Glucose and pH Sensing in Interstitial Fluids.

MinWook Park, Kyuil Chung, Young-Seong Kim, Jeon Woong Kang, Joong Yeon Lim

Department of Mechanical, Robotics, and Energy Engineering, Dongguk University, Seoul04620, Republic of Korea. · Laser Biomedical Research Center, G. R. Harrison Spectroscopy Laboratory, Massachusetts Institute of Technology, Cambridge, Massachusetts02139, United States.

Preclinical

Key points

From the paper's conclusion
  • Here, we present an integrated, capillary-driven hollow microneedle (HMN) patch that combines passive fluid collection, nitrocellulose (NC) filtration, and surface-enhanced Raman scattering (SERS) sensing of glucose and pH.

Abstract

Minimally invasive monitoring of interstitial fluid (ISF) biomarkers requires reliable sample collection and molecular analysis without external pumping. Here, we present an integrated, capillary-driven hollow microneedle (HMN) patch that combines passive fluid collection, nitrocellulose (NC) filtration, and surface-enhanced Raman scattering (SERS) sensing of glucose and pH. A three-dimensionally printed hydrophilic HMN structure enables skin access and capillary transport, while a flexible and optically transparent substrate permits Raman interrogation from the backside. The SERS substrate was fabricated by Langmuir-Blodgett assembly of four layers of silver-coated gold nanorods (GNR@Ag) on polydimethylsiloxane to create dense plasmonic hotspots. The substrate exhibited good spatial uniformity (RSD = 8.4%) and interbatch reproducibility (RSD = 4.8%). Functionalization with 4-mercaptophenylboronic acid and 4-mercaptobenzoic acid generated glucose- and pH-responsive signals, respectively. Glucose-dependent SERS responses were calibrated up to 30 mM (R2 = 0.97), and pH sensing was linear across pH 4-8 (R2 = 0.98). The NC filter reduced protein-related interference, and low-power excitation (0.25 mW) maintained spectral stability for 200 s. Crosstalk experiments revealed a pronounced pH dependence in glucose sensing but comparatively limited glucose interference with the pH readout. In mice, the patch exhibited fluid uptake after a 20 min collection period and glucose-dose-dependent SERS responses following intravenous administration. Ex vivo tests using excised mouse skin conditioned on agarose media at different pH values further demonstrated pH-responsive sensing under tissue-contact conditions. These results support the potential of a capillary-driven SERS HMN platform for minimally invasive glucose and pH sensing in ISF.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Capillary-Driven Surface-Enhanced Raman Scattering Hollow Microneedle Patch for Glucose and pH Sensing in Interstitial Fluids — Science Skim