Cavity-Enhanced Beat Frequency Light-Induced Thermoelastic Spectroscopy Using Differential-Frequency Demodulation

Hongqiang Fan, Mengpeng Hu, Hui Zhang, Dongqing Zhang, Qiang Wang

原始摘要(英文原文)· Original abstract
The calibration of the resonant frequency and Q factor of a quartz tuning fork (QTF) in light-induced thermoelastic spectroscopy (LITES) is essential, but time-consuming and disruptive to ongoing measurements. We report on a beat-frequency (BF) method based on differential-frequency demodulation to achieve precise and real-time calibration of QTF resonant frequency and Q factor while maintaining the full excitation of the QTF in LITES. This distinctive property leads to a double improvement in response amplitude compared to the well-optimized LITES operated with the widely used differential-frequency modulation. The tight locking of a near-infrared laser to a high finesse (∼12000) optical cavity further enhances the laser–gas interaction, and then the system’s sensitivity. By targeting the R(4) transition of hydrogen sulfide (H 2 S), we achieved a minimum detection limit of 157 parts per billion in concentration and a normalized noise equivalent absorption coefficient of 1.25 × 10 –12 cm –1 ·W·Hz –1/2 . Comparison with state-of-the-art LITES sensors confirms the superior performance achieved in this work.