ACS omega

Cationic Phenylpyridine Labeling for Improved Oligosaccharide and Glycan Analysis by Capillary Electrophoresis with LED-Induced Fluorescence Detection.

Filip Dusa, Michal Gregus, Richard Cmelik, Hubert Vesely, Hana Pizova, Pavel Bobal, Jana Lavicka

Institute of Analytical Chemistry of the Czech Academy of Sciences, Veveri 97, 602 00 Brno, Czech Republic. · Masaryk University, Faculty of Pharmacy, Department of Chemical Drugs, Studentska 832/13, 625 00 Brno, Czech Republic.

Key points

From the paper's conclusion
  • To fully exploit its fluorescence characteristics, we also optimized the capillary electrophoresis system by modifying the LED-induced fluorescence (LED-IF) detector through systematic evaluation of excitation LEDs and emission filters, resulting in substantially improved detection sensitivity.

Abstract

Capillary electrophoresis with laser-induced fluorescence detection (CE/LIF) is a powerful technique for analyzing glycans. However, it requires derivatization with a fluorescent compound to enhance both electrophoretic mobility and detection sensitivity simultaneously. The widely used labeling reagent 8-aminopyrene-1,3,6-trisulfonic acid (APTS) suffers from several limitations, including relatively low labeling efficiency and poor compatibility with mass spectrometry (MS), highlighting the need for improved alternatives. In this work, we designed and synthesized a novel cationic fluorescent label based on a phenylpyridine scaffold incorporating hydrazide functionality for efficient oligosaccharide and glycan derivatization via hydrazone formation. To fully exploit its fluorescence characteristics, we also optimized the capillary electrophoresis system by modifying the LED-induced fluorescence (LED-IF) detector through systematic evaluation of excitation LEDs and emission filters, resulting in substantially improved detection sensitivity. The optimized labeling, separation, and detection conditions enabled sensitive analysis of released glycans from various glycoproteins with detection limits in the nanomolar concentration range. The combination of this new phenylpyridine-based fluorescent label with the enhanced CE/LED-IF detection platform significantly improves the analytical performance of CE for glycan analysis and provides a more sensitive, robust, and versatile approach for comprehensive oligosaccharide and glycan characterization.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers