From the paper's conclusion
  • To our knowledge, this G/coumarin-benzothiazole/PLA composite is the first 3D-printable, organic-modified conductive material demonstrating superior electrochemical performance, offering a robust, scalable platform for portable serotonin monitoring in bioanalytical and clinical applications.

Abstract

3D printing enables low-cost, high-performance electrochemical sensors for bioanalytical and clinical use. Herein, we report, for the first time, a graphite/polylactic acid (G/PLA) conductive composite modified with a coumarin-benzothiazole derivative for 3D-printed electrodes targeting enhanced serotonin detection. The optimized composition (G/coumarin-benzothiazole/PLA, 35:5:60 wt %) was processed with a 3D pen to produce conductive electrodes without electrochemical activation. Electrochemical impedance spectroscopy showed an ≈74% decrease in charge-transfer resistance for the modified material, while cyclic voltammetry revealed an ≈137% increase in serotonin oxidation current and a distinct oxidation peak at +0.375 V vs Ag|AgCl|KCl(sat.). Optimized differential pulse voltammetry (modulation amplitude 70 mV; scan rate 30 mV s-1; step potential 8 mV) provided a linear range of 0.1-6.0 μmol L-1 (R 2 = 0.997), LOD = 0.075 μmol L-1, LOQ = 0.10 μmol L-1, and precision RSD < 5%. The sensor exhibited selectivity against common interferents (glucose, urea, uric acid, tyrosine, epinephrine, and dopamine), delivered near-quantitative recoveries in fortified human serum and cerebrospinal fluid, and interelectrode reproducibility RSD < 3%. To our knowledge, this G/coumarin-benzothiazole/PLA composite is the first 3D-printable, organic-modified conductive material demonstrating superior electrochemical performance, offering a robust, scalable platform for portable serotonin monitoring in bioanalytical and clinical applications.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers