Homemade Filament Incorporating an Organic Modifier (Coumarin-Benzothiazole) for the Fabrication of High-Performance Voltammetric Sensors for Serotonin Determination.
Natália M Caldas, Raíssa G Gouvea, Yuri P V de Carvalho, Amanda G Batista, Mikaelly O B de Sousa, Celia M Ronconi, Lucas V de Faria, Diego P Rocha,
Department of Analytical Chemistry, Institute of Chemistry, Fluminense Federal University, Niterói 24020-141, RJ, Brazil. · Department of Organic Chemistry, Institute of Chemistry, Fluminense Federal University, Niterói 24020-141, RJ, Brazil.
Key points
From the paper's conclusion- To our knowledge, this G/coumarin-benzothiazole/PLA composite is the first 3D-printable, organic-modified conductive material demonstrating superior electrochemical performance, offering a robust, scalable platform for portable serotonin monitoring in bioanalytical and clinical applications.
Abstract
3D printing enables low-cost, high-performance electrochemical sensors for bioanalytical and clinical use. Herein, we report, for the first time, a graphite/polylactic acid (G/PLA) conductive composite modified with a coumarin-benzothiazole derivative for 3D-printed electrodes targeting enhanced serotonin detection. The optimized composition (G/coumarin-benzothiazole/PLA, 35:5:60 wt %) was processed with a 3D pen to produce conductive electrodes without electrochemical activation. Electrochemical impedance spectroscopy showed an ≈74% decrease in charge-transfer resistance for the modified material, while cyclic voltammetry revealed an ≈137% increase in serotonin oxidation current and a distinct oxidation peak at +0.375 V vs Ag|AgCl|KCl(sat.). Optimized differential pulse voltammetry (modulation amplitude 70 mV; scan rate 30 mV s-1; step potential 8 mV) provided a linear range of 0.1-6.0 μmol L-1 (R 2 = 0.997), LOD = 0.075 μmol L-1, LOQ = 0.10 μmol L-1, and precision RSD < 5%. The sensor exhibited selectivity against common interferents (glucose, urea, uric acid, tyrosine, epinephrine, and dopamine), delivered near-quantitative recoveries in fortified human serum and cerebrospinal fluid, and interelectrode reproducibility RSD < 3%. To our knowledge, this G/coumarin-benzothiazole/PLA composite is the first 3D-printable, organic-modified conductive material demonstrating superior electrochemical performance, offering a robust, scalable platform for portable serotonin monitoring in bioanalytical and clinical applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An advanced carbon/bismuth oxide nanoparticle-based filament for multifunctional 3D-printed electrochemical sensors toward paracetamol and lead in pharmaceutical, clinical, and environmental electroanalysis.Mikrochimica acta · 2026
- A hybrid composite based on an imidazolate zeolite framework, graphite, and polylactic acid, housed in a 3D-printed electrode body, for the sensitive determination of the Sunset Yellow dye in beverage samples.Talanta · 2026
- Plasma-Activated 3D-Printed Electrode for the Determination of Tolfenamic Acid in Environmental Samples.ACS omega · 2026
- From design to detection: a review of 3d printing materials, electrochemical sensors, and biosensors.Anais da Academia Brasileira de Ciencias · 2026
- A Comparative Study of Conductive 3D Printing Filaments for Electrochemical Sensing Applications Pretreated by Alumina Polishing, Electrochemical Activation, and Electrodeposition of Au NanoparticlesACS electrochemistry. · 2025
- Embedded Printing-Derived Pyrolytic Carbon Electrodes with Graphene-like Features for Neural Stem Cell Differentiation and Dopamine SensingbioRxiv · 2026