A Hyperbranched Rolling Circle Amplification Assay Using Aptamer-Hemin/G-Quadruplex Recognition for the Detection of Creatine Kinase-MB.
Ensiyeh Mirsadoughi, Sayantan Tripathy, Fatemeh Zare, Gerard L Cote, Samuel Mabbott
Department of Biomedical Engineering, Texas A&M University, 101 Bizzell Street, College Station, Texas 77843, United States. · Center for Remote Health Technologies & Systems, Texas A&M Engineering Experiment Station, 600 Discovery Drive, College Station, Texas 77840-3006, United States.
Key points
From the paper's conclusion- Here, we report a sensitive colorimetric assay for CK-MB based on a conformation-locked aptamer guanine-quadruplex (G-quadruplex) system integrated with magnetic nanoparticle capture and hyperbranched rolling circle amplification (HRCA).
Abstract
Creatine kinase-myocardial band (CK-MB) remains an important biomarker for the early diagnosis and clinical management of myocardial infarction. Here, we report a sensitive colorimetric assay for CK-MB based on a conformation-locked aptamer guanine-quadruplex (G-quadruplex) system integrated with magnetic nanoparticle capture and hyperbranched rolling circle amplification (HRCA). Target binding induces conformational unlocking of the aptamer, initiating HRCA-mediated signal amplification and generating a concentration-dependent colorimetric response in the presence of a chromogenic substrate. A compact 3D-printed imaging module, combined with a Python-based processing pipeline, was used to extract grayscale intensity values from assay images for quantitative analysis. Using only 25 μL of sample, the assay achieved limits of detection (LOD) of 0.268 ng/mL in aqueous samples and 0.328 ng/mL in blood-spiked samples, and limits of quantification (LOQ) of 0.89 ng/mL and 1.09 ng/mL, respectively, with a linear dynamic range of 0.5-100 ng/mL. Grayscale intensity was inversely proportional to CK-MB concentration, enabling quantitative detection across clinically relevant levels. The HRCA amplification process and target-induced activation were further validated by TapeStation gel electrophoresis, confirming the formation of amplified DNA nanostructures. This platform provides a low-volume, amplification-based strategy for CK-MB detection that is compatible with simple optical readout and has strong potential for point-of-care applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Highly sensitive and specific high-throughput detection of Aβ42 in clinical specimens using a G-quadruplex-structured fluorescent aptamer analysis platform.Talanta · 2026
- Homogeneous Extraction-Free Dual-ctDNA Detection via DNA Nanomaterial Fusion for Rapid Breast Cancer DiagnosisACS Nano · 2025
- Quantitative detection of the diabetic retinopathy-associated microRNA-1281 via a CRISPR/Cas12a-RCA ratiometric sensing platform.Talanta · 2026
- A fluorescent aptamer-based method for kanamycin detection incorporating G-quadruplex DNAzyme and exonuclease III-assisted signal amplification.Chemical communications (Cambridge, England) · 2026
- A signal-locked clamped hybridization chain reaction nanoamplifier for rapid and highly sensitive analysis of Aβ42.Talanta · 2026
- Cryo-Engineered DNA Cryogels with Interconnected Macroporous Architectures for Enhanced Mass Transfer and Ultrasensitive cTnI Detection via an Integrated Microfluidic Platform.ACS sensors · 2026