RiboGreen-Based Encapsulation Efficiency Measurement for Lipid Nanoparticles: Evaluation of Total RNA Assay and Surrogate-Matrix Correction of Free RNA Fluorescence Bias
Yun Zhu, Zhe Tan, Harmon Johnson, Dylan Charland, Jiang Wu
Key points
From the paper's conclusion- The RiboGreen fluorescence assay is widely utilized to determine mRNA encapsulation efficiency (EE %) in lipid nanoparticle (LNP) formulations by separately quantifying free and total mRNA, but both measurements often exhibit high variability and inconsistent performance, compromising accurate EE % measurement.
Abstract
The RiboGreen fluorescence assay is widely utilized to determine mRNA encapsulation efficiency (EE %) in lipid nanoparticle (LNP) formulations by separately quantifying free and total mRNA, but both measurements often exhibit high variability and inconsistent performance, compromising accurate EE % measurement. We evaluated each assay using both LNP and targeted lipid nanoparticles (tLNP) formulations to identify root causes of poor performance. The total RNA assay performed acceptably, with %CV below 13% and inter-run recovery of 92−113% as assessed using nanoparticles with known mRNA concentration and different ionizable lipids, confirming complete RNA−lipid dissociation under Triton X-100 lysis. Progressive fluorescent signal decay was observed in the free RNA assay during prolonged incubation, highlighting the importance of an optimal time window for plate reading. More strikingly, the free assay yielded a systematic underestimation of over 30% when calibrated in conventional Trizma-EDTA (TE) buffer, inflating apparent EE % irrespective of dilution factor. Parallelism testing revealed that the calibration curve in TE buffer failed to match the responses in the sample matrix, attributable to nanoparticle-mediated fluorescence attenuation. Substituting the TE calibration buffer with empty nanoparticles (LNPs without mRNA) as a surrogate matrix restored inter-day spike-in recovery to 84−102% across 1−10% spike levels with %CV ≤ 18% at the lower limit of quantitation (LOQ), enabling accurate EE % reporting up to 99%. Matrix-matched calibration is therefore essential for reliable EE % determination and represents a broadly applicable solution for method qualification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Colloidal Light Scattering Distorts Fluorometric Quantification of mRNA in Lipid Nanoparticles.Current issues in molecular biology · 2026
- An Integrating Sphere Based Method for Quantifying RNA Encapsulation Efficiency in Lipid Nanoparticles without Lysis.Analytical chemistry · 2026
- Impact of cell line and reporter gene selection on in-vitro transfection evaluation of mRNA lipid nanoparticlesAAPS Open · 2025
- Single-Particle Multiparametric Microscopy Reveals Structural, Size, and Payload Heterogeneity in mRNA-Loaded Lipid NanoparticlesACS Nano · 2025
- siRNA/mRNA payload of lipid nanoparticles quantified using fluorescence microscopy and correlation spectroscopybioRxiv · 2026
- Deep-UV Resonance Raman Spectroscopy Enables Direct Quantification of mRNA in Intact Lipid Nanoparticle FormulationsAnalytical Chemistry · 2026