Investigations into Gas-Phase Formation in Glass CRMs during LA-ICP-MS
Sarah Julia Oberholzer, Barbara Umfahrer, Detlef Günther
Key points
From the paper's conclusion- These findings demonstrate that gas-forming elements can exhibit markedly different behavior within the ICP, which is important for the required internal standardization in LA-ICP-MS.
Abstract
Gas-phase formation during Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry (LA-ICP-MS) has so far only been observed and investigated in organic matrices, such as gelatine or polymers. In this study, pulse-resolved LA-ICP-TOFMS was used to investigate gas-phase formation during laser ablation of the glass reference materials NIST SRM 610 and NIST SRM 612. Temporal separation of the particulate and gaseous signal fractions using an elongated aerosol transport tube revealed measurable gaseous transport for Zn, Ge, As, Cd, Sn, and Sb, whereas all remaining investigated elements were transported exclusively in the particulate phase. The extent of gaseous transport depended on the glass reference material, laser fluence, and ablation mode, with consistently higher gaseous fractions observed for NIST SRM 612 than for NIST SRM 610. In contrast to previous observations for gelatine, the gaseous fractions of the glass matrices decreased with increasing laser fluence and were most pronounced during the beginning of the laser ablation process. The influence of gas-phase formation on quantitative analysis was evaluated by varying the internal standard during quantification of NIST SRM 612 using NIST SRM 610 as the calibration standard. Zn, Ge, and Cd were most accurately quantified using internal standards with similar gas-phase formation and aerosol transport behavior, whereas Sn, As, and Sb behaved similarly to the particle-transported matrix elements despite exhibiting measurable gaseous fractions. These findings demonstrate that gas-forming elements can exhibit markedly different behavior within the ICP, which is important for the required internal standardization in LA-ICP-MS. However, the presented results show that the suitability of an element as an internal standard cannot be predicted solely from the presence of a measurable gaseous fraction. Instead, gas-phase formation should be considered as an additional factor when selecting internal standards for quantitative LA-ICP-MS.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Method Development and Validation for Simultaneous In Situ Fe Isotope and Multi-Element Analysis by LASS MC-ICP-MS/ICP-ToF-MS.Analytical chemistry · 2026
- Matrix-dependent dissolution of CO2-laser-generated surrogate nuclear explosion debris.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Initial Performance Evaluation of an Ultraviolet 157 nm Laser Ablation System for ICP-MS Analysis.Analytical chemistry · 2026
- A Methodological Framework for Dual-Energy Synchrotron XFM Characterization of Stratified Historical Glass: A Peranakan Bead Case StudyACS Omega · 2026
- Process Regime Mapping of Green Picosecond Laser Delamination Through Glass in POE Photovoltaic Laminates After Field AgingApplied Sciences · 2026
- Photothermal Response and Multi-Factor Durability of Glass-Assisted Laser-Blackened Superhydrophobic Al-Mg SurfacesCoatings · 2026