Recovery of Quantitative Secondary Ion Mass Spectrometry under External Perturbations Relevant to In Situ and Operando Measurements.
Paweł Piotr Michałowski
Łukasiewicz Research Network - Institute of Microelectronics and Photonics, Warsaw02-668, Poland.
Key points
From the paper's conclusion- These results demonstrate that perturbation-specific calibration and extraction optimization can restore reproducible quantitative SIMS response under selected externally imposed perturbations, providing a metrological basis for the further development of quantitative in situ and operando SIMS.
Abstract
In situ and operando characterization techniques are becoming increasingly important for investigating electronic materials and devices. Although secondary ion mass spectrometry (SIMS) offers unmatched sensitivity, quantitative in situ and operando SIMS remains largely unexplored because external perturbations disturb ion optical conditions and quantitative calibration. Here, representative magnetic, electrical, and optical perturbations were investigated, and a practical strategy for recovering quantitative SIMS was demonstrated. Conventional restoration of the primary and secondary ion optics failed to provide reproducible dopant-to-matrix intensity ratios, yielding coefficients of variation between 26.2% and 54.6%. Introducing individually optimized extraction parameters reduced this variability to 0.3-1.1%. Long-term reproducibility was evaluated using 866 measurements of boron- and phosphorus-implanted reference materials acquired over approximately 6.5 years. Despite repeated source replacements and maintenance procedures, the calibrated dose and projected range remained reproducible with coefficients of variation below 1%. Finally, quantitative validation was performed using implanted reference materials measured under all investigated perturbations. Although the raw SIMS profiles differed substantially in both signal intensity and apparent depth scale, perturbation-specific calibration resulted in closely overlapping dopant distributions. These results demonstrate that perturbation-specific calibration and extraction optimization can restore reproducible quantitative SIMS response under selected externally imposed perturbations, providing a metrological basis for the further development of quantitative in situ and operando SIMS.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Critical Evaluation of Concentration Proxies in SIMS Diffusion StudiesAnalytical Chemistry · 2026
- PACE-SIMS: Checkpoint-Gated Autonomous SIMS Characterization with AI-Agent Quality ControlarXiv · 2026
- Decomposition of Secondary-Ion Kinetic Energy Distributions into Three Characteristic Components in MeV-SIMS.Journal of the American Society for Mass Spectrometry · 2026
- Quantitative Interleaved-Pulse Acquisition Extends the Range of Isotope-Ratio Measurements in ToF-SIMSarXiv · 2026
- Sample-Handling Bias and Direct ICP-MS of Ultratrace Metals in Semiconductor-Grade Isopropyl AlcoholACS Omega · 2026
- Extending the operating window of scanning electron microscopy through an integrated electron-optical architecture for high-temperature and near-ambient-pressure environments.The Review of scientific instruments · 2026