Key Laboratory of Tropical Biological Resources of Ministry of Education and Hainan Engineering Research Center for Drug Screening and Evaluation, School of Pharmaceutical Sciences, Hainan University, Haikou570228, China. · Research Center for Drug Metabolism, School of Life Sciences, State Key Laboratory of Supramolecular Structure and Materials, Center for Supramolecular Chemical Biology, College of Chemistry, Jilin University, Changchun130012, China.
Key points
From the paper's conclusion- Here we report a molecular ruler-based mass spectrometry strategy that integrates MSALL acquisition, PEG standard calibration anchors spanning broad molecular-weight ranges, and a self-built high-resolution mass spectral (HRMS) library to convert broad, overlapping PEG signal envelopes into temporally traceable, molecular-weight-referenced analytical targets.
Abstract
A growing number of long-acting protein therapeutics rely on high-molecular-weight (HMW) polyethylene glycol (PEG) conjugation to extend circulation time, yet the in vivo fate of PEG-derived species remains a critical blind spot for drug safety and regulatory evaluation. A key analytical challenge is that intact conjugates and PEG-derived species cannot be resolved by intact-protein or total-PEG measurements alone. Here we report a molecular ruler-based mass spectrometry strategy that integrates MSALL acquisition, PEG standard calibration anchors spanning broad molecular-weight ranges, and a self-built high-resolution mass spectral (HRMS) library to convert broad, overlapping PEG signal envelopes into temporally traceable, molecular-weight-referenced analytical targets. Applied to 40 kDa PEG-modified recombinant human growth hormone (PEG-rhGH) in rats, the workflow quantitatively traced the intact conjugate (plasma half-life t1/2 = 26.83 h) and concurrently resolved tissue-associated PEG 40 kDa and a PEG-related metabolite series. Intact PEG-rhGH remained below the validated tissue LLOQ, whereas PEG-derived components accumulated preferentially in the liver (peak at 48 h), with total urinary and fecal recovery of the quantitatively monitored intact PEG-rhGH and PEG 40 kDa below 10% by 144 hindicating profoundly delayed elimination. HRMS interrogation revealed low-molecular-weight PEG oligomers (PEG7-PEG21) together with carboxylic-acid- and aldehyde-terminated species in liver, consistent with in vivo chain shortening and end-group transformation. By linking intact-conjugate exposure, tissue accumulation, metabolite profiling, and excretion within a single transferable framework, this strategy provides a measurement-science tool for the systematic evaluation of HMW PEGylated protein therapeutics and other pharmaceutical polymer systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Anti-PEG antibodies in drug delivery: From origins to mitigation strategies.Advanced drug delivery reviews · 2026
- A Simplified Workflow for Robust, High-Throughput and Quantitative Plasma Proteomics through Streamlined In-Tip Proteolysis.Analytical chemistry · 2026
- Differential Ester Hydrolysis Enables Spatiotemporally Coordinated Cancer Combination Therapy Validated by Label-Free LC-MS/MS.Nano letters · 2026
- Comprehensive Biophysical and Biochemical Characterization of Multi-Arm Maleimide-Poly(ethylene glycol) Conjugated Bovine and Human Hemoglobin as Potential Red Blood Cell Substitutes.ACS applied bio materials · 2026
- A sensitive assay for estimation of DM1 fortified with trastuzumab in human serum utilizing positive LC-MS/MS electrospray ionization and its application to clinical studyBioanalysis · 2026
- Identification and quantitation strategies of circulating cysteine-conjugated linker-payload catabolites and free payload from a cleavable-linker ADC.Journal of pharmaceutical and biomedical analysis · 2026