SANDWiCH: An Oil-Free High-Throughput Microwell-Array Platform for Digital Loop-Mediated Isothermal Amplification.
Jeonghwan Oh, Dong Hyun Han, Jongho Park, Juhyun Lee, Kyung Tae Noh, Je-Kyun Park
Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology, Daejeon 34141, Republic of Korea. · Department of Infectious Disease Research, Armed Forces Medical Research Institute, Daejeon34059, Republic of Korea.
Key points
From the paper's conclusion- Here, we present serially-assembled nanodrop detection with a chip (SANDWiCH), an oil-free, high-throughput microwell-array digital LAMP platform designed for contamination-resistant multi-sample analysis.
Abstract
Digital nucleic acid amplification tests enable absolute molecular quantification by partitioning samples into independent reactors, but scalable digital loop-mediated isothermal amplification (LAMP) remains limited by oil-based compartmentalization, droplet stability requirements, and assembly-induced sample spreading in open microwell arrays. Here, we present serially-assembled nanodrop detection with a chip (SANDWiCH), an oil-free, high-throughput microwell-array digital LAMP platform designed for contamination-resistant multi-sample analysis. SANDWiCH integrates a 3 × 8 array layout, micropillar-assisted capillary loading, pressure-sensitive adhesive film sealing, and alignment-guided serial assembly to confine each sample to its assigned array before compartmentalization. This design enables spontaneous partitioning of reaction mixtures into 10,000 microwells, each with a volume of 7.85 pL, per array without pumps, actuators, or oil overlays, allowing 24 samples to be processed in parallel under identical amplification conditions. By varying the number of microwell-array units assigned per sample, the platform provides a tunable balance between assay throughput and quantification accuracy; the limit of quantification was reduced from 30.2 copies/μL in the highest-throughput mode to 1.3 copies/μL in the highest-accuracy mode. Using Plasmodium vivax as a target, quantitative analysis showed excellent agreement with theoretical predictions based on the Poisson distribution across six orders of magnitude (Pearson correlation coefficient, r = 0.991). Clinical validation with 60 blood-derived nucleic acid samples yielded 100% (95% CI, 88.4%-100%) sensitivity and 100% (95% CI, 88.4%-100%) specificity, while requiring only 0.29 μL of reagent per sample.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Droplet-Free and Oil-Free Microfluidics Based on Self-Pumping Hydrogel for Point-of-Care Multiplex Digital Nucleic Acid DetectionAnalytical Chemistry · 2026
- Single-Particle ICP-MS Enabled Multiplex Digital Detection of Nucleic Acids via Feedback Recombinase Polymerase Amplification.Analytical chemistry · 2026
- ThermiQuantTM MegaScan: high-throughput isothermal reactor with quantitative colorimetric readout for paper-based nucleic acid amplification testsnpj Biosensing · 2026
- AI-powered high-throughput digital colony picker platform for sorting microbial strains by multi-modal phenotypesNature Communications · 2025
- B-058 Validation of Droplet Digital PCR for High-Throughput Molecular QuantificationClinical Chemistry · 2026
- Temperature‐Resolved Coding for Highly Multiplexed Digital PCR Using MeltArraySmall · 2026