Cell-Surface-to-Solution Signal Transduction via a CRISPR/Cas12a-Based Dual-Target Aptamer Sensor for Detecting Urothelial Carcinoma Cells in Urine.
Shun-Li Wu, Jia-Yi Shi, Yan Tan, Hong-Xin Jiang, Xiao-Yun Hou, Shuang-Qing Liu, Kai-Di Li, Wen-Bo Sun,
State Key Laboratory of Medicinal Chemical Biology, Tianjin Key Laboratory of Biosensing and Molecular Recognition, Research Centre for Analytical Sciences, College of Chemistry, Nankai University, Tianjin300071, P. R. China. · Department of Urology, The Second Hospital of Tianjin Medical University, Tianjin Institute of Urology, Tianjin300211, P. R. China.
Key points
From the paper's conclusion- This work provides a capture-retain-activate strategy for the non-invasive analysis of intact exfoliated UC cells and illustrates a potential design route for coupling cell-surface recognition with solution-phase CRISPR/Cas12a signal amplification.
Abstract
Urothelial carcinoma (UC) is a common malignancy of the genitourinary system with high recurrence rates, requiring non-invasive postoperative monitoring. Direct detection of exfoliated UC cells in urine can provide tumor-origin evidence but remains challenging because of low target-cell abundance, complex urine interference, and subjective cytological interpretation. Herein, we developed a CRISPR/Cas12a-based dual-target aptamer sensor for sensitive and specific detection of exfoliated UC cells in urine. The sensing system spatially couples CKAP4-mediated magnetic capture with PD-L1-dependent retention of a Cas12a activator on the same captured cell. Magnetic washing removes unbound activator probes, whereas the cell-retained activator subsequently initiates Cas12a trans-cleavage in the solution-phase reaction. This capture-retain-activate architecture provides an AND-gate-like route for converting spatially coupled cell-surface recognition events into an amplified fluorescence output. The sensor achieved a detection limit of 12 cells per assay and showed good selectivity against interfering cells. A pilot clinical evaluation using 32 urine samples distinguished UC patients from non-UC individuals with an area under the receiver operating characteristic curve (AUC) of 0.996. This work provides a capture-retain-activate strategy for the non-invasive analysis of intact exfoliated UC cells and illustrates a potential design route for coupling cell-surface recognition with solution-phase CRISPR/Cas12a signal amplification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Modular Cell-Surface-to-Solution CRISPR-Cas12a Platform for Wash-Free Quantification of Cell-Surface Proteins on Living Cells.Analytical chemistry · 2026
- CRISPR/Cas12a cis -Cleavage-Mediated Smartphone-Controlled Light-Addressable Photoelectrochemical Biosensor for Highly Specific Point-of-Care Detection of Multiple Cancer BiomarkersAnalytical Chemistry · 2026
- Fluidly Confined CRISPR-Magnetic Microbots Empowered Homogeneous Electrochemical Biosensor for Amplified Detection and Discrimination of Cancer-Derived Extracellular Vesicle SubtypesAnalytical Chemistry · 2026
- A Novel Noninvasive Detection Method for Urinary Exfoliated Cells in Urothelial Carcinoma Based on Pressure Signals Mediated by Nanoparticles.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026
- Photoactivated liquid-liquid phase separation-based cascade CRISPR/Cas12a electrochemical platform for sensitive small extracellular vesicles detection.Talanta · 2026
- Split Activator Assembly via Aptamer Binding Enables CRISPR‐Cas12a Activation With Non‐Nucleic Acid TargetsSmall Methods · 2026