科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ Analytical chemistry2026-09-08

Novel Impingement-Plate Mouthpieces: Enhancing Data Quality in Breathomics Analysis via Direct High-Resolution Mass Spectrometry.

Huiling Wang, Dingyi Wang, Xin Luo, Dan Lei, Runqi Wang, Guibin Jiang, Keda Zhang, Xiaolan Hu, Xue Li

原始摘要(英文原文)· Original abstract
Breath sampling mouthpieces are essential for breathomics research, as they effectively remove saliva contamination during sampling. However, for real-time online mass spectrometric breath analysis, the widely adopted filter-based mouthpieces substantially compromise data quality due to analyte adsorption and background interference. In this study, guided by the inertial impaction principle, we designed and fabricated three impingement-plate mouthpiece types via 3D printing using biocompatible, low-porosity poly(lactic-co-glycolic acid). Computational fluid dynamics simulations and salivary amylase validation experiments confirmed efficient saliva droplet interception by the proposed configurations. Targeted detection of a ten-component gas standard mixture by secondary electrospray ionization high-resolution mass spectrometry (SESI-HRMS) demonstrated that impingement-plate types exhibited far fewer statistically significant deviations from the control group compared with the filter-based mouthpiece. Among the three designs, the reverse-raked configuration yielded responses closest to the control for most analytes. The types also provided improved signal stability and reproducibility, with interunit coefficients of variation (CVs, n = 10) of 6.43-29.84%, compared with 28.91-66.36% for the filter-based mouthpiece. Principal component analysis based on high-purity nitrogen purging tests showed that the impingement-plate mouthpiece groups clustered closer to the control group than the filter group, suggesting minimal background interference. Untargeted breath profiling confirmed that mass spectral features acquired with impingement-plate mouthpieces were highly consistent with those obtained from filter-free sampling, while the filter-based mouthpiece induced deviations in breath metabolic signatures. Overall, inertial impaction-based designs mitigate the trade-off between saliva droplet removal and analyte transmission, reducing mouthpiece-derived bias and improving data fidelity and analytical robustness in real-time breathomics.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Novel Impingement-Plate Mouthpieces: Enhancing Data Quality in Breathomics Analysis via Direct High-Resolution Mass Spectrometry. — 科研速览 Science Skim