Energy-Decoupled Photoelectrochemical and Pressure Dual-Mode Biosensors for Machine-Learning-Assisted Immunoassay.
Jinxin Liu, Zhen Yang, Xingxing Meng, Guohua Qi, Chuanping Li
Anhui Laboratory of Functional Coordinated Complexes for Materials Chemistry and Application, School of Chemical and Environmental Engineering, Anhui Polytechnic University, Wuhu241000, P. R. China. · State Key Laboratory of Electroanalytical Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun130022, P. R. China.
Key points
From the paper's conclusion- 3D-printed portable devices were further fabricated for signal acquisition, and a machine learning (ML)-assisted data analysis was employed for multivariate data fusion.
Abstract
Reliable and ultrasensitive detection of cardiac troponin I (cTnI) remains challenging due to the signal interference and limited self-validation capability of conventional biosensors. Herein, we report an orthogonal dual-mode biosensing strategy that integrates photoelectrochemical (PEC) sensing with catalysis-induced pressure transduction (CIPT) to achieve energy-decoupled signal generation and data-level cross-validation. Specifically, a multifunctional plasmonic Z-scheme probe, CdS@CdIn2S4/AuPt, was first synthesized to trigger in situ assembly of ZnO-based dual Z-scheme heterostructures upon cTnI recognition. This configuration simultaneously amplifies the photocurrent via enhanced charge separation and catalyzes the hydrolysis of ammonia borane to produce a macroscopic pressure change. These two outputs originate from distinct energy conversion pathways─interfacial charge transfer versus bulk gas expansion─thereby establishing an energy-decoupled dual-signal system. 3D-printed portable devices were further fabricated for signal acquisition, and a machine learning (ML)-assisted data analysis was employed for multivariate data fusion. The smart sensing platform exhibits high sensitivity for cTnI with limits of detection of 41.94 fg·mL-1 (PEC) and 164.80 fg·mL-1 (CIPT). The partial least-squares (PLS) model enables cross-channel validation, achieving a high prediction correlation (R2 > 0.9972) and improving recovery rates (98.74%-102.69%) in human serum. This work offers a robust and reliable pathway for the construction of energy-decoupled sensing platform for point-of care testing applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In Situ Engineering of S-Scheme Bi2O3/Zn3In2S6 Heterojunctions on Paper-Based Architectures for Dual-Mode Photoelectrochemical-Electrochemical Detection of Carcinoembryonic Antigen.Langmuir : the ACS journal of surfaces and colloids · 2026
- Self-Powered PEC Cathodic Immunosensor for CA125 Detection Based on the Cu₂ O Photocathode and the CdS/PANI PhotoanodeACS Sensors · 2025
- Single-Probe Dual-Mode Biosensor for Cross-Validated Detection of cTnI in Myocardial Damage Evaluation.ACS sensors · 2026
- Valence-Engineered Regulation of Built-In Electric Field in Hollow CuS@CeO2 Heterojunction for Photoelectrochemical Immunoassay.Analytical chemistry · 2026
- Ultrashort-Distance Internal Light Source-Integrated ECL-PEC Dual-Mode Molecular-Sensing Device for Highly Sensitive and Accurate DetectionAnalytical Chemistry · 2026
- Synchronous Visual-Photocurrent Dual-Readout for Cardiac Troponin I Sensing Enabled by Self-Powered Point-of-Care Platform.Analytical chemistry · 2026