Not One Method to Rule Them All: A Comparative Study of Chromatographic Platforms (RP-LC-, HILIC-, SFC-, and IC-HRMS) for Water Analysis
Jonathan Zweigle, Michael P. Schlüsener, Joana Flottmann, Tobias Bader, Nanna Hjort Vidkjær, Ulla E. Bollmann, Jan H. Christensen, Selina Tisler
University of Copenhagen · Federal Institute of Hydrology
Key points
From the paper's conclusion- We compared 12 chromatographic methods across four platforms, RP-LC, anion chromatography (IC), supercritical fluid chromatography (SFC), and hydrophilic interaction chromatography (HILIC), using 127 environmentally relevant compounds (log D pH7.4 −5.6 to 6.6).
Abstract
High Resolution Image Download MS PowerPoint Slide Reversed-phase liquid chromatography coupled to high-resolution mass spectrometry (RP-LC-HRMS) is the standard for nontarget screening (NTS) of environmental samples but lacks retention of highly polar contaminants. We compared 12 chromatographic methods across four platforms, RP-LC, anion chromatography (IC), supercritical fluid chromatography (SFC), and hydrophilic interaction chromatography (HILIC), using 127 environmentally relevant compounds (log D pH7.4 −5.6 to 6.6). Compounds were analyzed in solvent and for a polar subset, in groundwater enriched by vacuum evaporation. Data were collected across four laboratories using 5 RP-LC-, 3 HILIC-, 2 SFC-, and 2 IC-HRMS setups. Feature detection with standard tools yielded more false negatives for SFC and IC. To enable a fair and method-agnostic comparison, an extracted ion chromatogram (EIC)-based workflow was used. Of the 127 compounds, 125 were detected by at least one platform. For log D pH7.4 > 0, RP-LC covered ∼90%, followed by SFC (∼70%), while IC and HILIC each covered <30%. For very polar compounds (log D pH7.4 < 0), coverage dropped across all platforms. SFC and HILIC detected up to 60% of polar analytes; IC performed better in negative ionization mode, consistent with anion-exchange separation. Detection frequency declined with polarity, reflecting analytical limitations. Combining RP-LC with either SFC or HILIC increased coverage to 94%. In spiked groundwater, coverage was lower (73%) due to matrix effects and losses during vacuum enrichment. Peak widths were narrowest for SFC (∼2.5 s) and RP-LC (∼4 s) and broadest for HILIC (∼7 s) and IC (∼17 s). Retention times showed limited cross-platform correlation while ionization efficiency was consistent, except for SFC. As no single method provided full coverage, combining RP-LC with one complementary platform (SFC, HILIC, or IC) is required to extend chemical space in environmental NTS.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hyphenation of supercritical fluid chromatography to high resolution Orbitrap mass spectrometry for the purpose of non-targeted small molecule analysis.Journal of chromatography. A · 2026
- Comparing Suspect and Nontargeted Screening Workflows for Perinatal Chemical Exposome Analysis: A European Interlaboratory Study from the PARC-EU Project.Environmental science & technology · 2026
- Rapid Broad‐Polarity Screening of 459 Emerging Contaminants in Environmental Waters by Online Polarity‐Partitioned Two‐Dimensional Liquid Chromatography‐Tandem Mass SpectrometryJournal of Separation Science · 2026
- Trends in LC/GC‐HRMS Non‐Targeted Screening of Unknown Contaminants of Emerging Concern in Aquatic Systems: Workflows, No‐Standard Identification, and Environmental ApplicationsMass Spectrometry Reviews · 2026
- Uncovering monitoring gaps and novel persistent and mobile substances (PMs) in groundwater using lyophilisation enrichment and SFC-HRMS smart screeningAnalytical and Bioanalytical Chemistry · 2026
- Large-Scale Quantification of Xenobiotics Following Non-Targeted Liquid Chromatography Tandem Mass Spectrometry.Journal of the American Society for Mass Spectrometry · 2026