Two-center comparison of dedicated and general-purpose treatment planning for stereotactic body radiotherapy of spinal metastases with simultaneous integrated boost.

Marcel Nachbar, Sylvia Garny, Sebastian Hubertus Maier, Benjamin Tengler, Johannes Mücke, Vanessa da Silva Mendes, Daniel Felix Fleischmann, Sebastian Norbert Marschner,

Section for Biomedical Physics, Department of Radiation Oncology, University Hospital and Medical Faculty, Eberhard Karls University Tübingen, Tübingen, Germany. Electronic address: Marcel.nachbar@med.uni-tuebingen.de. · Department of Radiation Oncology, LMU University Hospital, LMU Medizin, LMU Munich, Munich, Germany.

Key points

From the paper's conclusion
  • TPSded achieved comparable target coverage, improved OAR sparing, and lower surrounding-tissue dose across both TPSgen plan sets. It also required the lowest median MU, suggesting potential advantages in plan quality and delivery for complex SIB-based spine SBRT.

Abstract

PURPOSE: Stereotactic body radiation therapy (SBRT) for spinal metastases requires steep dose gradients near organs at risk (OARs). Dedicated treatment planning systems may improve plan quality, but previous comparisons with general-purpose systems have yielded mixed results. This study compared plans from a dedicated spinal SBRT-optimized system with two independently generated plan sets from a general-purpose treatment planning system for SIB treatment of spinal metastases. METHODS: Twenty patients with 22 spinal target volumes were analyzed. Plans were generated with the same general-purpose system by two medical physicists at different centers (TPSgen1 and TPSgen2) and compared with plans from a dedicated site-specific system (TPSded). Dose-volume histogram parameters for targets, OARs, and a conformality structure were evaluated. Monitor units (MU), arc number, and cumulative gantry span were assessed. RESULTS: No significant differences were observed between TPSded and either TPSgen plan set for PTV D98% or D2%. For OARs and the conformality structure, 8 of 30 DVH parameters differed significantly between TPSded and TPSgen1, with 7 favoring TPSded, while 8 parameters differed significantly between TPSded and TPSgen2, all favoring TPSded. TPSded reduced spinal cord D2% by 2.7 Gy versus TPSgen1 and 2.6 Gy versus TPSgen2, and conformality-structure Dmean by 1.6 Gy and 2.6 Gy, respectively. TPSded required in median 247.0 MU/Gy, compared with 333.2 MU/Gy for TPSgen1 and 268.3 MU/Gy for TPSgen2. CONCLUSION: TPSded achieved comparable target coverage, improved OAR sparing, and lower surrounding-tissue dose across both TPSgen plan sets. It also required the lowest median MU, suggesting potential advantages in plan quality and delivery for complex SIB-based spine SBRT.