Image-processing-based detection of defective ultrasound transducer elements using clinical B-mode images.
Dingiswayo A Chikhwaza, Emily R Dirr, Adrianne R Kaiser-Vry, Stacie Madson-Whittaker, Timothy A Bigelow
Department of Electrical and Computer Engineering, Iowa State University, 2520 Osborn Dr, 2215 Coover Hall, Ames, 50011, IA, USA. · Department of Laboratory Animal Resources, Iowa State University, Ames, 50011, IA, USA.
Key points
From the paper's conclusion- Reliable artifact detection was achieved by averaging more than 50 images, resulting in a precise accuracy across all element conditions.
Abstract
The use of medical ultrasound is critical in healthcare, with applications ranging from medical diagnosis to therapeutic procedures. The elements of the ultrasound transducer are the core of this technology, and over time, they degrade and fail. Current maintenance practices rely on periodic maintenance, making them inefficient due to extended intervals between evaluations, which may lead to damaged transducers remaining in operation since visual identification of the defect is infeasible. This study presents a method that uses images collected by a transducer during normal workflow to determine a probe's status. Firstly, histogram equalization and frame averaging were applied to the images to reduce variability across clinical images while reinforcing consistent artifact patterns produced by inactive elements. The output then underwent a feature-extraction stage involving denoising and vertical edge filtering. To further distinguish the artifacts from the background, Otsu thresholding and morphological processing were done on the resulting image. Subsequently, component analysis based on geometric and spatial positioning was conducted to isolate true defects from residual noise. The identified feature was then overlaid on the original averaged input image to indicate the location of the artifacts. System validation was performed using linear B-mode images from of porcine subjects across different anatomical regions to simulate the variability encountered in clinical settings. The technique used a column-strength profile derived from Sobel-filtered images, in which artifacts produced threshold values greater than 1, unlike artifact-free regions. Reliable artifact detection was achieved by averaging more than 50 images, resulting in a precise accuracy across all element conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Detecting Uniformity Artifacts in Ultrasound Transducers: Insights from Clinical Median Images and Deep Learning for Automatic Detection.Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine · 2026
- Nonlinear mixing-wave imaging of additively manufactured specimens with different porosity levels using a multimode integrated transducer.Ultrasonics · 2026
- Encoder-free distortion correction for ultrasonic TFM C-scan images based on spatial mappingJournal of Physics: Conference Series · 2026
- Multi-system evaluation of baseline-referenced gray scale CNR profile deviation for phantom-based ultrasound quality assurance.Journal of applied clinical medical physics · 2026
- Complementary-Aperture Pulse Sequencing for Fundamental-Band Nonlinearity Parameter ImagingarXiv · 2026
- Defect detection using a laser diffuse ultrasonic phased array via selective transmitted wavesApplied Physics Letters · 2026